590614
37
Zoom out
Zoom in
Previous page
1/71
Next page
GMC-I Messtechnik GmbH 37
V/Hz, Ω, Temperature, and A/Hz Measurements
5.5 Temperature Measurement
Temperature measurement is performed with a type K
thermocouple (accessory, not included), which is connected to
the voltage input. Alternatively, a Pt100 or Pt1000 resistance
thermometer can be used with the METRAHITX-TRAOUTDOOR .
Selecting the Unit of Measure for Temperature
1nFo
... tEMP °C / °F 
(°C = default setting)
5.5.1 Measurement with Thermocouples, Temp TC
Ð Set the rotary switch to “Temp
TC
”.
Note
METRAHITX-TRAOUTDOOR only:
The last selected temperature measurement or the last
selected temperature sensor, i.e. type K or Pt100/Pt1000,
remains in memory and is accordingly displayed. Press the
FUNC | ENTER key in order to change to the other measuring
function if required.
Ð The reference temperature is measured at the internal
reference junction (see parameter “1tEMP ” in chapter 6.3
regarding querying).
Note
The internal reference temperature (temperature of the inter-
nal reference junction) is measured by a temperature sensor
inside of the instrument. This may be somewhat above room
temperature as a result of internal heat-up, or moving from
warmer to colder surroundings or vice versa.
Ð Connect the sensor to the two accessible jacks. The instru-
ment displays the measured temperature using the selected
unit of measure.
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
Temp
TC
0023.3
°C
Temp
RTD
Type K
TC
COM
V
Temp
Ω
TE (TC)
0256.0
°C
Pt1000
RTD
Measuring Range
TE (TC) K (NiCr-Ni) 250.0 +1372.0 °C
FUNC
ENTER
37


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE HIT at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE HIT in the language / languages: English as an attachment in your email.

The manual is 0,93 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt SECULIFE HIT

Gossen Metrawatt SECULIFE HIT User Manual - Dutch - 70 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info