590614
9
Zoom out
Zoom in
Previous page
1/71
Next page
GMC-I Messtechnik GmbH 9
Safety Precautions
Be prepared for the occurrence of unexpected voltages at
devices under test (e.g. defective devices). For example,
capacitors may be dangerously charged.
Make certain that the measurement cables are in flawless
condition, e.g. no damage to insulation, no interruptions in
cables or plugs etc.
No measurements may be made with this instrument in
electrical circuits with corona discharge (high-voltage).
Special care is required when measurements are made in HF
electrical circuits. Dangerous pulsating voltages may be
present.
Measurements under moist ambient conditions are not
permitted.
Be absolutely certain that the measuring ranges are not
overloaded beyond their allowable capacities. Limit values are
included in chapter 8, “Technical Data”, in the table entitled
“Measuring Functions and Measuring Ranges” in the “Over-
load Capacity” column.
The multimeter may only be operated with installed batteries or
rechargeable batteries. Dangerous currents and voltages are
otherwise not indicated, and the instrument may be damaged.
The instrument may not be operated if the fuse cover or the
battery compartment lid has been removed, or if its housing is
open.
The input for the current measuring range is equipped with a
fuse link (exception: METRAHITB
ASE).
Maximum permissible voltage for the measuring circuit (= rated
voltage of the fuse) is 1000 V AC/DC.
Use specified fuses only (see page 59)! The fuse must have a
breaking capacity of at least 30 kA.
Opening of Equipment / Repair
The equipment may be opened only by authorized service per-
sonnel to ensure the safe and correct operation of the equipment
and to keep the warranty valid.
Even original spare parts may be installed only by authorized serv-
ice personnel.
In case the equipment was opened by unauthorized personnel,
no warranty regarding personal safety, measurement accuracy,
conformity with applicable safety measures or any consequential
damage is granted by the manufacturer.
Repair and Parts Replacement
When the instrument is opened, voltage conducting parts may be
exposed. The instrument must be disconnected from the measur-
ing circuit before the performance of repairs or the replacement of
parts. If repair of a live open instrument is required, it may only be
carried out by trained personnel who are familiar with the dangers
involved.
Defects and Extraordinary Strains
If it may be assumed that the instrument can no longer be
operated safely, it must be removed from service and secured
against unintentional use.
Safe operation can no longer be relied upon:
If the device demonstrates visible damage
If the instrument no longer functions, or if malfunctioning
occurs
After long periods of storage under unfavorable conditions,
e.g. humidity, dust or extreme temperature (see “Ambient Con-
ditions” on page 60)
9


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE HIT at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE HIT in the language / languages: English as an attachment in your email.

The manual is 0,93 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt SECULIFE HIT

Gossen Metrawatt SECULIFE HIT User Manual - Dutch - 70 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info