SECULIFE ESPRO
INPUT ZERO – The Input Zero offset can be slightly different between RF donuts. This parameter
accesses an auto-zeroing function that eliminates this offset. Independent settings are saved for the
0.1:1 RF donut and 1:1 RF donut. The user can switch between the two donut types without having to re-
zero the input. The input needs to be zeroed only when a new donut is introduced.
INPUT RANGE
– The input range can be scaled to accommodate the signal that is being measured. The
input can be set to fixed ranges of 100mV Peak, 1000mV peak or Auto-Ranging. For Auto-Ranging
mode, the low range will be used for readings from 0.00 to 30.00 mV RMS. The high range will be used
for 20.0 to 700.0 mV RMS.
INPUT MODE
– There are two input modes to allow for measurement of continuous signals or pulsed
signals. In Continuous Mode, the input is updated every 100 mS. This mode should be used for all
Electrosurgical Generator waveform outputs in general use generators. A number of Electrosurgical
generator manufacturers offer generators with pulsed outputs, where there is a long duty cycle (typically
½-second or more) and the actual RF output of the generator is active for a brief period of time within the
duty cycle (typically 1/10
th
second or less). In Pulsed Mode, the input to the SECULIFE ESPRO is only
processed when a signal over 20 mV in amplitude is detected. The
SECULIFE ESPRO analyzes the
pulsed input waveform and can provide the RMS readings for either the overall input or the pulse only
(see Diagram 1, Page 15).
NOTE: When set to Pulsed
Mode, a small graphic appears in
the upper right corner of the
screen to identify to the user that
the
SECULIFE ESPRO is
looking for a pulsed RF input.
22 GMC-I Messtechnik GmbH
22


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ES PRO at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ES PRO in the language / languages: English as an attachment in your email.

The manual is 0,89 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info