GMC-I Messtechnik GmbH 81
Database and Entry Operations — LCD Pictographs
I
N
/ I
F
Z
L-N
/ Z
L-PE
EXTRA t
A
+I
EXTRA RCM
Saving a measured value with differing electrical circuit parameter
The electrical circuit parameter you have set at the test instrument does
not correspond to the parameter saved in the structure under object data.
Example: The residual operating current defined in the database is 10 mA,
whereas you have performed measurements with 100 mA. If you wish to
perform all future measurements with 100 mA, the value must be modi-
fied in the database by acknowledging it with . The measured value
will be documented and the new parameter will be accepted.
If you wish to leave the parameter in the database unchanged, press key
. Measured value and modified parameter will only be documented.
All Please enter a designation (alphanumeric).
All
Operation with a Barcode Scanner
Error message appears when the “EDIT” entry field is opened and battery
voltage is less than 8 V. Output voltage is generally switched off during
barcode scanner operation if U is less than 8 V in order to assure that
remaining battery capacity is adequate for entering designations for
devices under test and saving the measurement.
Remedy: Rechargeable batteries must be recharged, or replaced towards
the end of their service life.
All
Operation with a Barcode Scanner
Current flowing through the RS 232 port is too high.
Remedy:
The connected device is not suitable for this port.
All
Operation with a Barcode Scanner
Barcode not recognized, incorrect syntax
All
Data cannot be entered at this location within the structure.
Remedy: Observe profile for preselected PC software (see SETUP menu).
All
Measured value cannot be saved at this location within the structure.
Remedy: Make sure that you have selected the right profile for you PC
evaluation program in the SETUP menu (see section 4.6).
All
Memory is full.
Remedy: Save your measurement data to a PC and then clear memory at
test instrument by deleting the database or by importing an empty data-
base.
All
Delete measurement or database.
This prompt window asks you to confirm deletion.
SETUP
Data loss after changing language or profile,
or after restoring default settings.
Back up your measurement data to a PC before pressing the respective
key.
This prompt window asks you to confirm deletion.
All
This error message appears when the database, i. e. the structure cre-
ated in the ETC software, is too large for the device memory.
The database in the device memory is empty after database transfer has
been interrupted.
Remedy: Reduce the database in ETC or send the database without
measured vales (key Send structure) if measured values should already be
available.
Status
Test
plug
Meas.
Adapter
Position of the
Function Switch
Function / Meaning
81


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt PROFITEST MPRO at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt PROFITEST MPRO in the language / languages: English as an attachment in your email.

The manual is 2,28 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt PROFITEST MPRO

Gossen Metrawatt PROFITEST MPRO User Manual - Dutch - 92 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info