48 GMC-I Messtechnik GmbH
Database and Entry Operations — LCD Pictographs
Err38
IΔN / IF
ZL-N / ZL-PE
EXTRA tA+IΔ
Measured Value Storage with Deviating Electrical Circuit Parameter
The electrical circuit parameter selected by yourself at the test instrument does not
coincide with the parameter entered under object data in the structure.
Example: Residual operating current is specified as 10 mA in the database, but you
have performed measurement with 100 mA. If you want to perform all future mea-
surements with 100 mA, the value in the database has to be changed by acknowl-
edging with the key. The measured value is documented and the new param-
eter is accepted.
If you want to leave the parameter in the database unchanged, press the key.
The measured value and the changed parameter are only documented in this case.
Err39 All (page 39) Please enter a designation (alphanumeric).
Err40 All
Operation with a Barcode Scanner
Error message when the “EDIT” entry field is opened and rechargeable battery
voltage is less than 8.0 V. Output voltage is generally switched off during barcode
scanner operation if U is less than 8.0 V, in order to assure that remaining battery
capacity is adequate for entering designations for devices under test and saving
the measurement.
Remedy: rechargeable batteries must be recharged, or batteries must be replaced
towards the end of their service life.
Err41 All
Operation with a Barcode Scanner
Current flowing through the RS 232 port is too high.
Remedy:
the connected device is not suitable for this port.
Err42 All Operation with a Barcode Scanner
Barcode not recognized, incorrect syntax
Err43 All Data cannot be entered at this location within the structure.
Remedy: observe profile for preselected PC software (see SETUP menu).
Err44 All
Measured value cannot be saved at this location within the structure.
Remedy: make sure that you have selected the right profile for you PC evaluation
program in the SETUP menu (see section 4.5).
Err45 All
Memory is full.
Remedy: save your measurement data to a PC and then clear memory at test
instrument by deleting the database or by importing an empty database.
Err46 All
Delete measurement or database.
This prompt window asks you to confirm deletion.
Err47 SETUP
Data loss after changing language or profile,
or after restoring default settings.
Back up your measurement data to a PC before pressing the respective key.
This prompt window asks you to confirm deletion.
Err48 All
This error message appears if the database, i.e. the structure created in ETC, is too
large for the instrument’s internal memory.
The database in the instrument’s internal memory is empty after database transfer
has been interrupted.
Remedy: reduce the size of the database in ETC or transfer the database without
measured values (Transmit Structure key), if measured values already exist.
Status
Error
No.
Position of the
Function Switch
Function / Meaning
48


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt Profitest Intro at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt Profitest Intro in the language / languages: English as an attachment in your email.

The manual is 1.69 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info