24 GMC-I Messtechnik GmbH
10 Characteristic Values
1)
Display range up to 600 V
2)
Display range up to 1.2 kV
3)
the indicated accuracy is only achieved with the shielded high-resistance measur-
ing cable KS-C (article no. Z541F)“
4)
does not conform to DIN EN 61557-2
5)
up to 5 Ω
Meas.
Qty.
UiSO Range Measuring Range
Reso-
lution
Open-Circuit
Voltage U
0max
Test
Current
Intrinsic Uncertainty
Measuring
Uncertainty
Overload
Capacity
RINS
BASE/TECH: 50 V
BASE/TECH: 100 V
INTRO: 250 V ... 1000 V
BASE: 250 V / 500 V
TECH: 250 V / 500 V / 1000 V
100 k 10.0 kΩ ... 99.9 kΩ 0.1 k
50 V /100 V:
1.25 U
ISO
250 V /
500 V /
1000 V:
1.1 U
ISO
I
N
= 1 mA
I
K
5 mA
±(5% rdg. + 3 d) ±(7% rdg. + 3 d)
METRISO BASE:
600 V AC/DC
TRMS
METRISO INTRO
METRISO TECH:
TRMS
1 M 100 kΩ ... 999 kΩ 1 k
10 M 1.00 MΩ ... 9.99 MΩ 10 k
100 M 10.0 MΩ ... 99.9 MΩ 100 k
1 G 100 MΩ ... 999 MΩ 1 M
10 G 1.00 GΩ ... 9.99 GΩ 10 M
100 G 10.0 GΩ ... 99.9 GΩ 100 M ±(8% rdg. + 3 d)
3)
±(10% rdg. + 3 d)
3)
200 G 100 GΩ ... 199 GΩ 1 G ±(25% rdg. + 5 d)
3)
±
(50% rdg. + 20 d)
3) 4)
U
AC/
DC
METRISO BASE
100 V 10.0 V ... 99.9 V 0.1 V
±(2.5% rdg. + 3 d) ±(5% rdg. + 3 d)
600 V AC/DC
TRMS
500 V 100 V ... 510 V
1)
1 V
METRISO INTRO
METRISO TECH
100 V 10.0 V ... 99.9 V 0.1 V
±(2.5% rdg. + 3 d) ±(5% rdg. + 3 d)
1000 V AC/DC
TRMS
1000 V 100 V ... 999 V
2)
1 V
RLO 10 Ω 0.17 ... 9.99 Ω
0.01
Ω
4 V < U0
< 6 V
200 mA I
I
260 mA
±(2.5% rdg. + 3 d) ±(5% rdg. + 3 d)
METRISO BASE:
600 V AC/DC
TRMS
METRISO INTRO
METRISO TECH:
1000 V AC/DC TRMS
R
METRISO BASE
METRISO TECH
Display range as of 01.0
Ω
100 Ω 10.0 ... 99.9 Ω 0.1 Ω
U
0
max. 15 V
1 mA I
I 1,3 mA
±(2.5% rdg. + 3 d) ±(5% rdg. + 3 d)
METRISO BASE:
600 V AC/DC
TRMS
METRISO TECH
1000 V AC/DC TRMS
1 kΩ 100 ... 999 Ω 1 Ω
10 kΩ 1.00 ... 9.99 kΩ 10 Ω
24


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRISO TECH at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRISO TECH in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRISO TECH

Gossen Metrawatt METRISO TECH User Manual - German - 6 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info