16 GMC-I Messtechnik GmbH
5 Measuring Insulation Resistance – Riso/Rins Function
5.1 Connection
Note
Checking the Measurement Cables
Before performing insulation measurement, the test
probes on the measurement cables should be short-cir-
cuited in order to assure that the instrument displays a
value very close to 0 Ω (see section 8). This makes it pos-
sible to detect interrupted measurement cables, which
simulate high insulation resistance.
For this instrument,
fuse testing
is only performed after swit-
ching from one insulation resistance measuring range to
another. When further measurements are performed in the
same measuring range, fuse testing is dispensed with for the
sake of a faster test procedure.
Ð Connect the device under test to the + and COM jacks.
Sample connection layouts for insulation resistance measurement are
included in section 14.1. Insulation resistance can only be measured at
voltage-free objects. If mains voltage or external voltage is applied to the
measurement inputs, measurement cannot be started. This is indicated
by the
high-voltage symbol
which appears at the display.
5.2 Performing the Measurement
Note: Condensation must be ruled out when performing measure-
ments at close to the freezing point.
Ð Select the measuring function and the desired test voltage
with the rotary switch, e.g. R
iSO
100V.
The momentarily selected test voltage (nominal voltage) appears
in the bottom right-hand corner of the display.
Ð
Start an individual measurement by briefly pressing the
START
key, or
initiate continuous measurement by briefly pressing the
CONTIN
key
.
L1
L2
N
PE
L3
R
B
R
E
+
COM
R
ISO
+ COM SHIELD
Rx
0V !
!
R
ISO
R
INS
MAN
DATA
ISO
500V
1
0
0
k
1
G
1
0
k
1
M
1
0
G
1
00
G
1
0
0
M
1
0
M
---
START
ENTER
16


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRISO TECH at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRISO TECH in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRISO TECH

Gossen Metrawatt METRISO TECH User Manual - German - 6 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info