590655
60
Zoom out
Zoom in
Previous page
1/68
Next page
60 GMC-I Messtechnik GmbH
The measurement has been saved to memory.
MERISO PRIME+(with feature B1 or B2)
The displayed high-voltage module has been detected. Turn device on with key switch.
The displayed high-voltage module has been detected. Turn device on with key switch.
Green signal lamps do not light up.
The displayed high-voltage module has not been detected.
– Mains cable not plugged in
– Key switch set to “0”
– Blown fuse
– Defective signal lamp
– Plug in the mains cable.
– Set the key switch to “I”.
– Replace the fuse
– Service
Red signal lamps do not light up.
– No data transfer between MERISO PRIME+ and 204HP/HV
– Defective signal lamp
– See “Device defective!” error message.
– Replace signal lamp, service
High-voltage module ready to be switched on
High-voltage is present at the test probes during measurement.
Do not touch the test probes!
The specified limit value I
MAX
has been exceeded. Current limiting has
switched the device to the “standby” mode.
Test completed successfully. Residual voltage less than 25 V
Phase angle display
Continuous operation with high test power during high voltage testing Allow the device to cool for approximately 10 minutes.
One of the two signal lamps in the high-voltage module is defective. Send the high-voltage module to GMC-I Service GmbH for
repair.
Displays and Error Messages Meaning / Cause Remedy
or
60


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRISO PRIME plus at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRISO PRIME plus in the language / languages: English as an attachment in your email.

The manual is 1,51 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRISO PRIME plus

Gossen Metrawatt METRISO PRIME plus User Manual - German - 68 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info