Measurement Types
You can choose one of two different types of measurement:
Measuring sequence with automatic polarity reversal (reversal
of current flow direction)
Manual measurement with specified current flow direction
8.1 Measurement with Automatic Polarity Reversal – AUTO Function
Automatic polarity reversal is selected as a default value. MAN
does not appear at the display.
Ð Start measurement in both current flow directions by briefly
pressing the START key for one-time polarity reversal or start
continuous measurement for permanent polarity reversal by
briefly pressing the CONTIN. key.
After the measuring sequence has been started, the instrument performs
measurement with automatic polarity reversal, first with current flow in
one direction (
Ω→
), and then in the other (
←Ω
).
DATA
blinks at the display
until the measured value has settled in. The larger (worst) measured value
is displayed.
←Ω→
appears at the display.
Difference > 10%
If, during measurement with automatic polarity reversal, the difference
between
Ω→
and
←Ω
is greater than 10%, the resistance values for
both polarities (current flow directions) are displayed separately (sepa-
rated by an underscore).
Δ
Rlo>10% appears at the display.
Limit Value Indication
If the measured value is less than or equal to 2 Ω, the Limit LED
lights up green. If the measured value is greater than 2 Ω, the LED
lights up red. As a prerequisite, the LiLED parameter must be set
to “on” (see page 13).
Resistances which do not demonstrate a stable value until after a
“settling in period” should not be measured with automatic polar-
ity reversal. Measurement with automatic polarity reversal may
lead to varying and/or inflated measurement values, and thus to
an ambiguous reading.
Ω→
←Ω
ΔRlo>10%:
←Ω & Ω
AUTO: Ω→
DATA
00.1
DATA
00.7
Rlo
07._0.1
Δ Rlo>10%
Ω→
Rlo ←Ω
DATA
Limit
< Limit R
LO
= OK 3
←Ω
ΔRlo 10% ? ←Ω→
ΔRlo > 10% ? Ω & Ω
21


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRISO INTRO at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRISO INTRO in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRISO INTRO

Gossen Metrawatt METRISO INTRO User Manual - German - 6 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info