590650
19
Zoom out
Zoom in
Previous page
1/28
Next page
GMC-I Messtechnik GmbH 19
7 Measuring Alternating Voltage
Sinusoidal alternating voltage within a frequency range of 40 to 200 Hz
can be measured with the test instrument.
Ð Press the U~/f key.
Ð Contact the measuring points with the measuring probes.
Note
+
The input impedance for the voltage measuring range is 5 M
8 Measuring Temperature and Humidity with Adapter Z541A
(Accessory Adapter)
Temperatures within a range of –10.0 C to +50.0 C, as well as relative
humidity from 10.0% to 90.0% can be measured with the temperature-
humidity combination adapter.
Ð Connect the temperature-humidity adapter to the IR interface of the
test instrument (see drawing on page 2) by inserting the guide pin of
the adapter in the opening provided at the top of the test instrument
such that the adapter is located on the two rubber pads in central
position to the housing. Then push the adapter down to fix it tightly.
Ð Press the T/F
REL
key. The adapter is activated via the interface.
Ð Select the desired temperature unit of measure, C or F, with the
corresponding key. Temperature and humidity are displayed directly.
Ð We recommend moving the test instrument with the adapter to and
fro for a few seconds to enable the adapter to adjust to the prevailing
room climate more quickly. The air current can thus penetrate through
the opening into the adapter housing more quickly. It would otherwise
take several minutes to achieve this condition.
Ð When changing to another function, the adapter is deactivated.
Ð The measured values are only updated every 5 seconds to spare the
adapter batteries.
Display at the Test Instrument
The „———“ symbol shown on the display can be attributed to the fol-
lowing causes:
incidence of sunlight
depleted batteries in adapter Z541A
adapter defective or not properly positioned
Adapter Z541A should not be exposed to intensive sunlight to prevent
inadvertent activation via infrared (battery life!).
19


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRISO C at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRISO C in the language / languages: English as an attachment in your email.

The manual is 0,77 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRISO C

Gossen Metrawatt METRISO C User Manual - Dutch - 28 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info