14 GMC-I Messtechnik GmbH
4 General Operation
The test leads are connected to the “+” and “COM” jacks.
Measuring High-Value Resistance with the KS-C Probe (Option)
When measuring electrostatic
discharge capacity for floor
coverings, the shielded cable
should also be connected to
the COM and SHIELD jacks (KS-C
accessory set, “cable set con-
sisting of measurement cable
and high-resistance measure-
ment cable for measurements
in the GΩ range”, see diagram). Be sure to observe color coding.
For measuring with the Test Probe for Remote Triggering (Option Z550A)
see section 14.5.1.
4.1 Switching On, Monitoring and Switching Off
If battery voltage falls below the allowable limit value (U < 8 V) the
instrument cannot be switched on, or it is immediately switched
off.
Measurement cannot be started in the resistance measuring
ranges in the event of external voltage.
The instrument only switches itself off automatically after comple-
tion of an automatic measuring sequence, and after the predeter-
mined on-time has expired (APOFF parameter in SETUP switch
position, see page 12). On-time is reset to its original value as
defined in the setup menu, as soon as any key or the rotary func-
tion switch is activated.
If the instrument is switched off automatically with the rotary
switch in any position other than OFF, it can be reactivated by
pressing the stby OFF key.
The instrument can be switched off manually by turning the rotary
switch to the OFF position.
Optical Indicators
1
Prerequisite: The LiLEd parameter is set to on (see page 13).
2
Function testing should be executed regularly (see following section on testing
LEDs).
+
C
O
M
S
H
I
E
L
D
KS-C
(Z541F)
LED Status Function – Cause
Limit value indication
1
Measured insulation resistance does not violate the limit
value.
– Measured low-resistance Rlo does not violate the limit
value.
Limit value indication
1
Measured insulation resistance has fallen short of the se-
lected limit value.
Measured low-resistance Rlo does has exceeded the per-
missible limit value.
External voltage when on or off (not M550N) and during discharge
2
Dangerous voltage of greater than 50 V is present at the mea-
surement inputs:
Initialization of the (insulation) resistance and
low-resistance measurements is disabled.
The discharging cycle has not yet been completed,
e.g. residual voltage at capacitive devices under test
Test/Measuring voltage with device switched on
Dangerous voltage of greater than 50 V is present at the mea-
surement inputs.
LCD Status Functions
Display
External voltage when off, and during discharge
Dangerous voltage of greater than 50 V is present at the
measurement inputs:
Initialization of the insulation resistance and
low-resistance measurements is disabled.
Test voltage applied during insulation measurement
The discharging cycle has not yet been completed,
e.g. residual voltage at capacitive devices under test
Limit
Green
Limit
Red
!
Red
14


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRISO BASE at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRISO BASE in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRISO BASE

Gossen Metrawatt METRISO BASE User Manual - German - 6 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info