4.4.1 Standard measurements 0 - 2 M
Tests with 200mA in accordance with EN/IEC 61557-4
can only be performed for resistance < 10 Ω and
must be performed with manual polarity change.
Press button
shortly to start the measure-
ment. With measurement values <10 the
METRAOHM 413 automatically switches to Lo-
mode and continues measuring with a
measurement current of 200 mA (tests comply
with EN/IEC 61557-4 ).
The display shows the symbol „Lo-“ and the
right LED briefly signalizes the 200 mA
measurement current.
If the measurement value is out of measurement
range (>1999 k) the symbol “OL“ (Overflow)
appears on the display. The measurment result is
displayed for approx. 3 s, after this the display
changes to „+ Rdy.
Change of polarity:
After measurement with positive polarity
(see above): press button
+/-
(“- Rdy“ and
green LED display readiness for measurement
with dierent polarity). Perform measurements
with negative polarity in the same way.
Large deviations between values are an indication
for application of galvanic voltage. Repeat both
measurements. If the results deviate from each
other in a similar way than before, an average value
between the “+“ and “-“ may be assumed.
4.4 Measuring
Attention!
Measuring point must be free of voltage!
Button
may only be pressed after establishing
contact to the measuring point if there is an absence
of voltage (see 4.5). Otherwise the device may be
damaged.
If necessary, clean the measurement point from
corrosion or paint.
Switch on the device with button
(+ Rdy and
the green LED signalize that the device is ready
for measurement).
Place the test electrode onto the measuring point
vertically if possible and assure safe contact.
If the red LED ashes and you hear an acoustic
warning signal, an external voltage is present!
The test must be interrupted (see 4.5).
Short warning signals can be caused by inductive
voltages or static charging which have no inuence
on the measurement.
Note: Direction of current ow is indicated in the
display as “+“ or “-“. The displayed sign refers to the
test probe.
23
23


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRAOHM 413 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRAOHM 413 in the language / languages: English, German, French as an attachment in your email.

The manual is 0,56 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info