590656
22
Zoom out
Zoom in
Previous page
1/48
Next page
22
4.3 Zero balancing
The METRAOHM 413 must be calibrated together
with the entire measurement line before each
measurement test series (see measurement setup
4.2). To take temperature changes into account, zero
balancing is also necessary if the same measurment
line is used again after a period of time.
Balancing occurs automatically for all measuring
ranges of 1μA ... 200 mA.
For balancing the device has to be switched o.
Contact the test electrode of the METRAOHM
413 to the reference earth and ensure safe
contact.
Press button
and then the button
at the
same time. The zero balance will be performed.
The display shows “CAL. After expiration of
time the value “0.00 ... 0.04 Ω“ is indicated on
the display.
Attention!
Maintain contact as long as “CAL“ is displayed
and the time expires.
If the message “Err“ appears, zero balancing has
failed and must be repeated. This error occurs if the
test probe slips during balancing or if the measure-
ment line is too high-resistive (>3.5 ).
For test purposes we recommend that the rst
measurement always be made directly at the
reference earth (see section 4.4). The result
should be < 0.04 for both polaritys.
Note: Long coiled measuring lines may cause faulty
measurments by their inductance. Resistance of the
measurement line remains in storage until a new
zero balance is performed. This value also remains in
storage if the measuring instrument is switched o,
or if the battery is replaced.
22


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRAOHM 413 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRAOHM 413 in the language / languages: English, German, French as an attachment in your email.

The manual is 0,56 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info