44 GMC-I Messtechnik GmbH
V/Hz, Ω, Temperature, and A/Hz Measurements
5.7.2 Alternating Current and Frequency Measurement,
Direct Connection, A AC and Hz
Ð First disconnect supply power from the measuring circuit or
the power consumer, and discharge any capacitors.
Ð In accordance with the current or frequency to be measured,
turn the rotary switch to A~ or Hz.
Ð Select the desired measured quantity by briefly pressing the
multifunction key FUNC | ENTER. Each time the key is pressed,
AC
TRMS
and Hz are alternately selected, and switching is ac-
knowledged with an acoustic signal.
Ð Safely connect the measuring instrument (without contact re-
sistance) in series to the power consumer as shown.
Ð Switch supply power to the measuring circuit back on.
Ð Read the display. Make a note of the measured value if the in-
strument is not being operated in the memory mode or the
transmission mode.
Ð Disconnect supply power from the measuring circuit or the
power consumer once again, and discharge any capacitors.
Ð Remove the test probes from the measuring point and return
the measuring circuit to its normal condition.
COM mA A
~
R
x
~
R
x
1
3
2
A~
Hz
A~
Hz
050.100
Hz
AC
!
I > 10 A
!
I > 16 A
Current measurement may only
be performed with installed bat-
teries!
Current
FUNC
ENTER
300 μA / 3 mA
3 A / 10 A (10 A max. 5 min)
Measuring Ranges:
30 mA / 300 mA
*
Additional temperature display
(internal temperature at jacks)
in the 3 A and 10 A ranges
03
.500
A
TRMS
24.6° C
*
Hz
A
(16 A max. 30 s)
8
.
8
.
8
.
8
.
8
03.500
A
AC
TRMS
24. C
*
015
.
000
Hz
44


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRAHIT ULTRA at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRAHIT ULTRA in the language / languages: English as an attachment in your email.

The manual is 0,9 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info