18 GMC-I Messtechnik GmbH
Control Functions
4 Control Functions
4.1 Selecting Measuring Functions and Measuring Ranges
4.1.1 Automatic Range Selection
The multimeter is equipped with auto-ranging for all measuring
functions, except for temperature measurement, and diode and
continuity testing. Auto-ranging is active as soon as the
instrument is switched on. The instrument automatically selects
the measuring range which allows for highest possible resolution
of the applied quantity. When the instrument is switched to
frequency measurement, the previously selected voltage
measuring range remains active.
AUTO-Range Function
The multimeter is switched automatically to the next higher range
at ±(3099 d + 1 d 0310 d), and to the next lower range at
±(280 d - 1 d 2799 d).
4.1.2 Manual Measuring Range Selection
Auto-ranging can be deactivated and measuring ranges can be
selected manually in accordance with the following table by
pressing the MAN / AUTO button.
The desired measuring range can then be selected with the
or
key.
The instrument is automatically returned to range selection when
the MAN / AUTO key is pressed, the rotary switch is activated or the
instrument is switched off and back on again.
Overview: Auto-Ranging and Manual Range Selection
* Via manual measuring range selection only
The multimeter is held in the selected measuring range. If the
range limit is exceeded, 0L appears at the display. You should then
switch to the next higher measuring range with the help of the
key.
Measurement type MΩ
ISO
: If the measured value is less than 10% of
the measuring range after manual measuring range selection, u r
(under range) appears at the display. You should then select the
next smaller measuring range with the help of the
key.
Function Display
MAN / AUTO
Manual mode active:
utilized measuring range is fixed
MAN
or
Range switching sequence for:
V: 300 mV*
3 V
30 V
300 V
600 V
Hz: 300 Hz
3 kHz
30 kHz
300 kHz (Hz(U))
Ω:
300
Ω ↔
3 k
Ω ↔
30 k
Ω ↔
300 k
Ω ↔
3 M
Ω ↔
30 M
Ω
A: 300 μA
3 mA
30 mA
300 mA
1 A
A : See chapter 5.9.3 und chapter 5.9.4
F: 30 nF 300 nF 3 μF 30 μF 300 μF
MΩ
ISO
:
300 k
Ω ↔
3 M
Ω ↔
30 M
Ω ↔
300 M
Ω
MAN
MAN / AUTO
Return to automatic measuring range selection
18


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRAHIT T-COM at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRAHIT T-COM in the language / languages: English as an attachment in your email.

The manual is 0,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRAHIT T-COM

Gossen Metrawatt METRAHIT T-COM Quick start guide - English, German - 2 pages

Gossen Metrawatt METRAHIT T-COM Quick start guide - Dutch - 3 pages

Gossen Metrawatt METRAHIT T-COM User Manual - Dutch - 72 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info