590608
64
Zoom out
Zoom in
Previous page
1/80
Next page
64 GMC-I Messtechnik GmbH
Technical Data
Insulation Resistance Measurement in Selector Switch Positions (a-b) (a-E) (b-E)
1)
1)
During insulation resistance measurement (
M
Ω
@UISO
): If
ERROR is displayed as
„Error“
>> limits:
U
interference
> 10 ... 20 V and U
interference
U
ISO
, Ri < 10 kΩ @ Uiso 10 V,
Ri < 100 kΩ @ Uiso 100 V
2)
Interference voltage measurement TRMS (V AC + DC) with 120 kΩ input resist-
ance, bandwidth 15 Hz ... 500 Hz, measuring error 3% + 30 Digit
Ripple voltage < 4 Vss 50 Hz, parallel capacitance < 3 μF
Interference-immune Capacitance Measurement in Selector Switch
Positions (a-b) (a-E) (b-E)
Measuring voltage U
0
= 2 Vss approx. 1 Hz, parallel resistance > 5 M
Ω
Ripple voltage < 4 Vss 50 Hz
Galvanic Signature Ascertainment in Selector Switch Positions (a-b)
Prerequisites:
Parallel resistance > 200 kΩ, parallel capacitance < 1 μF,
Ripple voltage < 4 Vss 50 Hz
and exclusively high-resistance DC voltage
Influencing Quantities and Influence Error
1)
With zero balancing
Measuring Range Resolution
Nominal Voltage
U
ISO
(U
INS
)
Intrinsic Error
under Reference
Conditions
±
(% rdg + d)
6 V ... 110 V
2)
0.1 V
Ri
approx.
120 k
Ω
3 + 30
5 ... 310.0 kΩ 0.1 kΩ 10//100 V 3 + 5
0.280 ... 3.100 MΩ 1 kΩ
10
//100 V
3 + 5
02.80 ... 31.00 MΩ 10 kΩ
10
//100 V
5 + 5
028.0 ... 310.0 MΩ 100 kΩ
10
//100 V
5 + 5
Measuring
Function
Switch
Setting
Nom.
Voltage
U
N
@100k
Open-
Circuit
Voltage
U
o
Nom.
Cur-
rent I
N
@100k
Short-
Circuit
Cur-
rent I
k
Acoustic
Signal
for
Overload Capacity
Value Time
U
interference
/
MΩ
ISO
— —— —
U > 110
V
110 V Cont.
MΩ
ISO
100 V Max.
120 V
> 1.0
mA
<1.2
mA
U > 110
V
100 V
10 s
8.7 V 11 V 0.09 < 0.260 U > 110V 100 V 10 s
Measuring
Range
Resolution
3100/1100
Digit
Intrinsic Uncertainty under
Reference Conditions
± ( ... % rdg. + ... d)
Measuring
Cycle (max.)
Overload
Capacity
300 nF 100 pF 2 + 10 2 s
600 V / PTC
max. 10 s
3 μF 1 nF 2 + 10 2 s
10 μF 10 nF 5 + 10 2 s
Signature if interference value is stable in range plus decision value
available
50% ... +30%
approx. +60% ... +200%
not available < +20%
Influencing
Quantity
Sphere of
Influence
Measured Quantity /
Measuring Range
1)
Influence Error
(...% rdg. + ... d) / 10 K
Temperature
0 °C ... +21° C
and
+25° C ... +40° C
V 0.2 + 5
V 0.4 + 5
300 Ω ... 3 MΩ 0.5 + 5
30 MΩ 1 + 5
mA/A 0.5 + 5
mA/A 0.8 + 5
30 nF ... 300 μF1 + 5
Hz 0.2 + 5
°C/°F (Pt100/Pt1000) 0.5 + 5
64


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRAHIT T-COM PLUS at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRAHIT T-COM PLUS in the language / languages: English as an attachment in your email.

The manual is 0,9 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRAHIT T-COM PLUS

Gossen Metrawatt METRAHIT T-COM PLUS Quick start guide - English, German - 2 pages

Gossen Metrawatt METRAHIT T-COM PLUS Quick start guide - Dutch - 2 pages

Gossen Metrawatt METRAHIT T-COM PLUS User Manual - Dutch - 80 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info