GMC-I Messtechnik GmbH 65
Technical Data
Power Supply
Battery 2 ea. 1.5 V mignon cell (2 ea. size AA),
alkaline manganese per IEC LR6
Service life With alkaline manganese batteries: approx.
200 hours (without M
@UISO
measurement)
Battery test Battery capacity display with battery
symbol in 4 segments: .
Querying of momentary battery voltage via
menu function.
Power OFF function
The multimeter is switched off automatically:
– If battery voltage drops to below approx.
1.8 V
– If none of the keys or the rotary switch are
activated for an adjustable duration of 10
to 59 minutes, and the multimeter is not in
the continuous operation mode
Power pack socket
If the power pack has been plugged into
the instrument, the installed batteries are
disconnected automatically.
Rechargeable batteries can only be
recharged externally.
1)
Times 0.7 for interface operation
Electrical Safety
Safety class II per EN 61010-1:2001/VDE 0411-1:2002
Measuring category CAT II CAT III
Nominal voltage 1000 V 600 V
Fouling factor 2
Test voltage 5.2 kV~ per EN 61010-1:2001/VDE 0411-
1:2002
Fuses
Fuse link FF 10 A/1000 V AC/DC;
10 mm x 38 mm; switching capacity: 30 kA
at 1000 V AC/DC;
protects the current measurement input in
the 300 A through 10 A ranges
Electromagnetic Compatibility (EMC)
Interference emission
EN 61326-1:2006, class B
Interference immunity
EN 61326-1:2006
EN 61326-2-1:2006
Measuring
Function
Nominal
Voltage U
N
Resistance
of the
DUT
Service life
in Hours
Number of Possible
Measurements with
Nominal Current per
VDE 0413
V 200
1)
V 150
1)
M
100 V 1 M 50
100 V 100 k 3000
500 V 500 k 600
1000 V 2 M 200
65


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRAHIT ISO at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRAHIT ISO in the language / languages: English as an attachment in your email.

The manual is 0,98 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRAHIT ISO

Gossen Metrawatt METRAHIT ISO Quick start guide - English, German - 2 pages

Gossen Metrawatt METRAHIT ISO Quick start guide - Dutch - 2 pages

Gossen Metrawatt METRAHIT ISO User Manual - Dutch - 72 pages

Gossen Metrawatt METRAHIT ISO Quick start guide - French - 2 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info