12 GMC-I Messtechnik GmbH
10 Temperature Measurement
10.1 Temperature Measurement with Pt100 and Pt1000
Ð Enter the type of sensor to be used (Pt100 or Pt1000) in the
menu mode:
rAtE SEnSor Pt 100 Pt 1000
Ð Set the selector switch to
°C
2
” for 2-wire measurement connection or
°C
4
” for 4-wire measurement connection.
Ð Connect the sensor as shown in the following diagram:
The instrument displays the measured temperature in the unit
of measure entered in the menu mode (parameter: “tEMP”).
Compensating for Cable Resistance and Offset
Due to high measuring resolution, compensation must be
made for offset and cable resistance in the order indicated
below, especially for the 2-wire resistance temperature
measuring function (“°C
2
):
Offset
Any remaining influence caused by the cables and contact
resistance can be eliminated by means of zero balancing (see
chapter 6.1, page 9).
Cable Resistance
Default setting: A preset cable resistance value is
compensated for when the default setting is used. The
preset value is 0.1 Ω which is suitable for temperature
sensors available as accessories.
Entering a value other than the preset cable resistance value:
(value from data sheet or user calculated, see below):
Enter the utilized sensor type (Pt100 or Pt1000) and the
cable resistance value in the “Setup” menu (range: 00.01
to 99.99 Ω):
rAtE SEnSor Pt 100 (Pt 1000)
Lr (lead resistance) XX.XX Ω 
Calculating cable resistance:
í Heat or cool the sensor to a known temperature (e.g.
0 °C in ice water), and adjust the value in the Lr menu
until the correct measured value is displayed.
4-Wire Connection2-Wire Connection
Possible Sensors:
Z3409
TF550
TF220
TS Chipset
ϑ
°
C
2
+
°
C
2
+
°
C
4
+
12


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRAHIT 30M at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRAHIT 30M in the language / languages: English as an attachment in your email.

The manual is 0,39 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info