6-12
Journal Interval for Timed Readings
Journal Interval for Timed Readings
Introduction
The Journal Interval screen allows users to define how often the Energy Platform saves
data periodically, regardless of power quality disturbances and journal limits.
Definitions
Power Values
Power Values Interval - Power parameter values are based upon all cycles during a one
second interval. Data is aggregated or summarized into min, max and avg values over
the averaging period and stored at the end of the interval. Data measured include watts,
power factors, as well as the one-second rms.
Aggregate Snapshot - Contains properties that measure periodic sample or snapshot of
the voltage and/or current waveform. Users are able to determine how often the
instrument will record waveform snapshots by turning the function on or off.
Demand and Energy
Demand Sub-Interval and Sub-Intervals per Interval - Refer to the time interval used in
calculating power demand values. Demand calculations are made every sub-interval on
the values that occurred during the most recent Demand Interval. Values set for
Demand Sub-Interval must be an integer-divisor of the Demand Interval since the
former applies when updating certain parameters of the latter.
Harmonics
Harmonics Journal Interval - Various harmonic parameters such as total harmonic
distortion (THD), total interharmonic distortion (TID), and telephone influence factor
(TIF) can be trended using periodic readings that are stored in the journal. Depending
on the harmonic journal interval set, the instrument records a sampling of the waveform
synchronized to the fundamental frequency, to ensure accurate calculation of harmonic
phase angles.
131


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt MAVOWATT 20 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt MAVOWATT 20 in the language / languages: English as an attachment in your email.

The manual is 2,89 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info