GMC-I Messtechnik GmbH 9
To test your ohmmeter, proceed as follows:
1 Connect the two probes of your measuring instrument to
the test sockets B1 and B2 (blue and black socket), if you
whish to select the 1 resistor.
2) Connect the two probes of your measuring instrument to
the test sockets B3 and B4 (blue and black socket), if you
whish to select the 50 m resistor.
3) On your ohmmeter, select the desired measuring range
according to the operating instructions of your ohmmeter.
4) Perform the resistance test.
5) Read the measured value on your measuring instrument.
6) Check that the displayed value corresponds to the value
selected on the ISO-Kalibrator 1, taking into account the
tolerances of the ohmmeter.
The resistance value at a time is given on the resistance
symbol which is correlated with the test sockets. You
should use this value as a comparison value. It was individ-
ually determined for your ISO-Kalibrator 1.
7) Conclude the measurement and put the ISO-Kalibrator 1
into an electrically dead condition.
8) You now can select another test resistor and start a new
test, or disconnect your measuring instrument from the
ISO-Kalibrator 1.
9) Do not on any account remove the probes in live condition.
9


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt ISO-KALIBRATOR 1 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt ISO-KALIBRATOR 1 in the language / languages: English as an attachment in your email.

The manual is 0,2 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info