21 ENGLISH
ment.
In order to get a more accurate measurement, it should be used in a quiet and comfortable environ-
ment.
When it is carried from cold environment to warm or humid environment, please do not use it imme-
diately.
If the device is splashed or coagulated by water, please stop operating.
DO NOT operate the device with sharp things.
High temperature, high pressure, gas sterilizing or immersion disinfection for the device is not per-
mitted. Refer to User Manual in the relative chapter (6.1) for cleaning and disinfection. Please turn off
the device before cleaning and disinfection.Please take out the internal battery before cleaning and
disinfection.
The device is suitable for children and adult.
The device may not be suitable for all users, if you can’t get a satisfactory result, please stop using
it.
Data averaging and signal processing have a delay in the upgrade of SpO2 data values. When
the data update period is less than 30 seconds, the time for obtaining dynamic average values will
increase, which is arisen from signal degradation, low perfusion or other interference, it depends on
the PR value.
The device has 3-year service life, date of manufacture: see the label.
This device has the function of prompting, users can check on this function according to chapter
5.5.1 as a reference.
The device has the function of limits prompting, when the measured data is beyond the highest or
lowest limit, the device would start prompting automatically on the premise of the prompting function
is on.
The device has the function of prompting, this function can either be paused, or closed (default set-
ting) for good. This function could be turned on through menu operation if you need. Please check
the chapter 5.5.1as a reference.
The device hasn’t low-voltage prompt function, it only shows the low-voltage, please change the
battery when the battery voltage is used up.
The maximum temperature at the SpO2 probe -tissue interface should be less than 41℃ which is
measured by the temperature tester.
During measuring, when abnormal conditions appear on the screen, please pull out your nger and
reinsert it to measure again.
If some unknown error appears during measuring, remove the battery to terminate operating.
Do not contort or drag the wire of the device.
The plethysmographic waveform is not normalized, as a signal inadequacy indicator, when it is not
smooth and stable, the accuracy of the measured value may degrade. When it tends to be smooth
and stable, the measured value read is the optimal and the waveform at this time is also the most
standard.
If necessary, please visit our ofcial website to get the information about SpO2 probe that can be
used with this device.
If the device or component is intended for single-use, then the repeated use of these parts will pose
risks on the parameters and technical parameters of the equipment known to the manufacturer.
If necessary, our company can provide some information (such as circuit diagrams, component lists,
illustrations, etc.), so that the qualied technical personnel of the user can repair the device compo-
nents designated by our company.
The measured results will be inuenced by the external colouring agent (such as nail polish, colour-
ing agent or color skin care products, etc.), so don’t use them on the test site.
As to the ngers which are too cold or too thin or whose ngernail is too long, it may affect the meas-
ured results, so please insert the thicker nger such as thumb or middle nger deeply enough into
the probe when measuring.
The nger should be placed correctly (see Attached gure 6), as improper installation or improper
contact position for sensor will inuence the measurement.
The light between the photoelectric receiving tube and the light-emitting tube of the device must
pass through the subject’s arteriole. Make sure the optical path is free from any optical obstacles like
rubberized fabric, to avoid inaccurate results.
Excessive ambient light may affect the measured results, such as surgical light (especially xenon
light sources), bilirubin lamp, uorescent lamp, infrared heater and direct sunlight, etc. In order to
4


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gima OXY 50 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gima OXY 50 in the language / languages: English as an attachment in your email.

The manual is 1.06 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gima OXY 50

Gima OXY 50 User Manual - French - 19 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info