712058
21
Zoom out
Zoom in
Previous page
1/112
Next page
EN
POSSIBLE PROBLEMS AND SOLUTIONS
Problem
SpO
2
or PR
cannot be shown
normally.
SpO
2
or PR is
shown
unstably
The oximeter
Cannot be
switched
On.
„Finger out“
Displayed on
screen
Indication lamps
suddenly
go off
“Error3” or
“Error4”
Displayed on
screen
“Error7”
displayed
on screen
Possible Reason
Finger is not
inserted correctly.
Patient’s oxyhaemoglo-
bin level is too low
to be measured
Finger might not be
inserted far enough.
Finger is trembling or
the patient’s body is
moving.
Battery power may be
inadequate or not be
there at all.
Batteries may have
been inserted incor-
rectly.
The oximeter might be
damaged.
There is no nger on
the pulse oximeter or
nger is not positioned
properly.
The product is automa-
tically switched off if no
signal is detected for
longer than 8 seconds.
Low power/ Batteries
empty.
Low power/ Batteries
empty
Receiving sensor
shielded or
sensor damaged.
Low power/ Batteries
empty
Infrared sensor
damaged.
Solution
Try inserting the nger
again.
Try several more times
if you are sure that there
is no problem with the
device. Please go to a
hospital in good
time to obtain a precise
diagnosis.
Try inserting the nger
again.
Try not to move.
Please replace batteries.
Please reinsert the
batteries.
Have the device tested by
the manufacturer
Place your nger in the
oximeter.
Normal
Replace the batteries.
Insert new batteries.
Have the device tested by
the manufacturer.
Change the batteries.
Have the device tested by
the manufacturer.
1.
2.
1.
2.
1.
2.
3.
1.
1.
2.
1.
2.
1.
2.
1.
2.
1.
2.
1.
2.
3.
1.
1.
2.
1.
2.
1.
2.
21


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Geratherm GT-300C203 - Oxy Control at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Geratherm GT-300C203 - Oxy Control in the language / languages: English, German, Dutch, French, Italian, Portuguese, Turkish, Spanish as an attachment in your email.

The manual is 20,73 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info