985
Users Manual
8
Operation
The subsequent sections tell you about Product
operation.
Power On/Off
Push to turn-on and turn-off the Product.
When the Product is turned on, a splash screen shows,
followed by the default sample screen.
You can start to collect samples immediately from the
sample screen. The samples use the default or stored
sample setup values that you can configure.
Before you use the Product, make sure the battery is fully
charged. See “The Battery” section.
Product Test
Electrical noise, sensor leakage, or other interference can
cause the Product to give incorrect data.
To check operation:
1. Attach the Zero-Count Filter.
2. In the Data Display Setup screen, enable the 0.3 μm
channel and set Concentration mode to COUNTS.
3. In the Sample Setup screen:
1. Set the sample time to 5 minutes
2. Set the hold time to 00:00:00
3. Set the delay time to 00:00:03
4. Set the MODE to Automatic
5. Set the number of CYCLES = 2.
4. Start the Product sampling and let it complete the 2 x
5 minute samples.
5. Examine the particle counts from the sample. Counts
must align with these specifications for instrument
operation to be verified:
No more than 1 particle >0.3 μm in 5 minutes
How to Purge the Product Sensor
Particles must be purged (removed) from the Product
sensor before use in a clean room or clean environment,
and after each high-sampling count. To purge the sensor:
1. Attach the adapter and the Zero-Count Filter.
2. Set the Product to sample continuously (set the
Count Mode to Cumulative.)
3. Start the count procedure. Continue the count
procedure until there are no new counts.
4. Remove the Zero-Count Filter for normal operation.
16


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 985 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 985 in the language / languages: English as an attachment in your email.

The manual is 0,72 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info