763942
86
Zoom out
Zoom in
Previous page
1/104
Next page
8845A/8846A
Users Manual
4-16
Function modifiers:
None
Checking Diodes
The diode function sends a current through a semiconductor junction while the Meter
measures the voltage drop across the junction (or junctions). Measurements are displayed
in the 10 V range at a relatively fast measurement rate. "OPEN" is displayed for voltages
10 % above the compliance voltage setting. Typical junction voltage drop on good
junctions are 0.3 to 0.8 volts. If enabled, the beeper will make a short beep when a good
junction is detected. Shorted diodes will indicate a significantly lower voltage.
With the higher compliance voltage (up to 10 V), the Meter’s diode test function can test
zeners up to 10 volts, diode stacks and LEDs. The selectable current and maximum
voltage allows you to tailor the diode test to the expected voltage for the junction being
tested.
To check a diode:
1. Press S twice on the 8846A or D once on the 8845A. An example of the diode
test display is shown below.
0.5994
F1 F2 F3 F4 F5
+
caw13f.eps
2. Select the test voltage and current appropriate for the diode being tested by pressing
the associated soft keys.
3. Connect the test leads as shown in Figure 4-8.
2
INPUT SENSE
HI
LO
FRONT
REAR
10 A
1V
00
mA
4
W
HI
LO
2W/4W
V
300V
FUSED
1000 V CAT I
600V CAT II
4
caw024.eps
Figure 4-8. Diode Testing Connections
86


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8846A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8846A in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8846A

Fluke 8846A User Manual - German - 108 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info