763942
23
Zoom out
Zoom in
Previous page
1/104
Next page
Introduction and Specifications
Electrical Specifications 1
1-13
Additional Crest Factor Errors (<100 Hz)............ Crest factor 1-2, 0.05 % of full scale
Crest factor 2-3, 0.2 % of full scale
Crest factor 3-4, 0.4 % of full scale
Crest factor 4-5, 0.5 % of full scale
Input Characteristics
Resolution
Range Resolution 4½ Digits 5½ Digits 6½ Digits Input Impedance
100 mV 100.0000 mV 10 μV 1 μV 100 nV
1 V 1.000000 V 100 μV 10 μV 1 μV
10 V 10.00000 V 1 mV 100 μV 10 μV
100 V 100.0000 V 10 mV 1 mV 100 μV
1000 V 1,000.000 V 100 mV 10 mV 1 mV
1 MΩ ±2 % shunted
by <100 pf
8846A Accuracy
Accuracy is given as ± (% measurement + % of range)
Range Frequency 24 Hour
(23 ±1 °C)
90 Days
(23 ±5 °C)
1 Year
(23 ±5 °C)
Temperature
Coefficient/ °C
Outside 18 to 28 °C
3 – 5 Hz 1.0 + 0.03 1.0 + 0.04 1.0 + 0.04 0.1 + 0.004
5 – 10 Hz 0.35 + 0.03 0.35 + 0.04 0.35 + 0.04 0.035 + 0.004
10 Hz – 20 kHz 0.04 + 0.03 0.05 + 0.04 0.06 + 0.04 0.005 + 0.004
20 – 50 kHz 0.1 + 0.05 0.11 + 0.05 0.12 + 0.05 0.011 + 0.005
50 – 100 kHz 0.55 + 0.08 0.6 + 0.08 0.6 + 0.08 0.06 + 0.008
100 mV
100 – 300 kHz [1] 4.0 + 0.50 4.0 + 0.50 4.0 + 0.50 0.20 + 0.02
3 – 5 Hz 1.0 + 0.02 1.0 + 0.03 1.0 + 0.03 0.1 + 0.003
5 – 10 Hz 0.35 + 0.02 0.35 + 0.03 0.35 + 0.03 0.035 + 0.003
10 Hz – 20 kHz 0.04 + 0.02 0.05 + 0.03 0.06 + 0.03 0.005 + 0.003
20 – 50 kHz 0.1 + 0.04 0.11 + 0.05 0.12 + 0.05 0.011 + 0.005
50 – 100 kHz 0.55 + 0.08 0.6 + 0.08 0.6 + 0.08 0.06 + 0.008
1 V
100 – 300 kHz [1] 4.0 + 0.50 4.0 + 0.50 4.0 + 0.50 0.2 + 0.02
3 – 5 Hz 1.0 + 0.02 1.0 + 0.03 1.0 + 0.03 0.1 + 0.003
5 – 10 Hz 0.35 + 0.02 0.35 + 0.03 0.35 + 0.03 0.035 + 0.003
10 Hz – 20 kHz 0.04 + 0.02 0.05 + 0.03 0.06 + 0.03 0.005 + 0.003
20 – 50 kHz 0.1 + 0.04 0.11 + 0.05 0.12 + 0.05 0.011 + 0.005
50 – 100 kHz 0.55 + 0.08 0.6 + 0.08 0.6 + 0.08 0.06 + 0.008
10 V
100 – 300 kHz [1] 4.0 + 0.50 4.0 + 0.50 4.0 + 0.50 0.2 + 0.02
3 – 5 Hz 1.0 + 0.02 1.0 + 0.03 1.0 + 0.03 0.1 + 0.003
5 – 10 Hz 0.35 + 0.02 0.35 + 0.03 0.35 + 0.03 0.035 + 0.003
10 Hz – 20 kHz 0.04 + 0.02 0.05 + 0.03 0.06 + 0.03 0.005 + 0.003
20 – 50 kHz 0.1 + 0.04 0.11 + 0.05 0.12 + 0.05 0.011 + 0.005
50 – 100 kHz 0.55 + 0.08 0.6 + 0.08 0.6 + 0.08 0.06 + 0.008
100 V
100 –- 300 kHz [1] 4.0 + 0.50 4.0 + 0.50 4.0 + 0.50 0.2 + 0.02
3 – 5 Hz 1.0 + 0.015 1.0 + 0.0225 1.0 + 0.0225 0.1 + 0.00225
5 – 10 Hz 0.35 + 0.015 0.35 + 0.0225 0.35 + 0.0225 0.035 + 0.00225
10 Hz – 20 kHz 0.04 + 0.015 0.05 + 0.0225 0.06 + 0.0225 0.005 + 0.00225
20 – 50 kHz 0.1 + 0.03 0.11 + 0.0375 0.12 + 0.0375 0.011 + 0.00375
50 – 100 kHz [2] 0.55 + 0.06 0.6 + 0.06 0.6 + 0.06 0.06 + 0.006
1000 V
100 – 300 kHz [1][2] 4.0 + 0.375 4.0 + 0.375 4.0 + 0.375 0.2 + 0.015
[1] Typically 30 % reading error at 1 MHz
[2] 1000 Volt range is limited to 8 X 107 volt-Hertz
23


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8846A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8846A in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8846A

Fluke 8846A User Manual - German - 108 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info