Front-Panel Operation
Configuring the Meter for a Measurement 3
3-9
The beeper state is stored in non-volatile memory and does not change when the Meter is
turned off or after a remote interface reset. The beeper is enabled when the Meter is
shipped from the factory.
Setting the Display Resolution
The steps for setting the Meter’s displayed resolution, varies depending on the selected
function. The dc volts, dc current and ohms functions set the resolution based on the
power line cycle (PLC) setting. AC volts, ac current, capacitance and temperature set
resolution using a low, medium or high setting.
To set the Meter’s display resolution dc volts, dc current and ohms:
1. Press L to see the measurement Setup menu.
2. Press the soft key under the RESOLUTION #DIG PLC label to reveal the
resolution selection menu.
The soft key labels are updated with five choices.
4 DIGIT .02 PLC
5 DIGIT .2 PLC
5 DIGIT 1 PLC
6 DIGIT 10 PLC
6 DIGIT 100 PLC
These choices determine the displayed resolution (4½, 5½, and 6½ digits) and the
measurement cycle time in reference to the power line frequency in cycles (PLC)
For example, selecting 5 DIGIT 1 PLC will display 5½ digit resolution and
take a measurement using one power line cycle A/D integrate period. For 60 Hz
power, a measurement is almost once every 60th of a second or 16.6666 milliseconds.
3. Press the soft key labeled with the desired resolution.
Note
Some of the faster PLC choices are not available when 2nd MEAS or DCV
Ratio are selected.
To set the Meter’s display resolution for ac volts, ac current, capacitance and
temperature:
1. Press L to see the measurement Setup menu.
2. Press the soft key under the RESOLUTN label to reveal the resolution selection
menu.
Three soft key labels are updated with HIGH, MEDIUM, and LOW. The actual
number of digits displayed will depend on the selected function and the Meter’s
range.
3. Press the soft key labeled with the desired resolution.
Setting the AC Signal Filter
There are three ac filter settings that can be employed to help make more accurate
measurements: 3 Hz Slow, 20 Hz, and 200 Hz.
For the ac volts and ac current functions, a filter selection is available as a modifier.
Pressing the Filter soft key will expose a menu that allows you to select between the
three filter settings.
53


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8845A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8845A in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8845A

Fluke 8845A User Manual - German - 108 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info