Operating the Meter from the Front Panel
Selecting a Measurement Function 3
3-13
HI
LO
1000V
750V
2W/4W
MAX
INPUT SENSE
HI
LO
4
W
10 A
mA
200 mA
MAX 10 A
MAX
1000V CAT I
600 V CAT II
500 V pk
V
1V
300V
Ground tab
eue26.eps
Figure 3-7. Input Connections for 4-Wire Ohms Using 2x4 Wire Leads
Measuring Current
W Caution
To avoid blowing the current fuse or damaging the Meter, do
not apply power to the circuit under test until test leads are
properly installed to the appropriate input terminals. For current
measurements above 200 mA, install the test leads into the 10 A
and LO terminals only.
The Meter is capable of making ac and dc current measurements up to 10 A.
To perform a current measurement:
1. Turn off power to the circuit under test.
2. Connect test leads between the Meter and the circuit under test.
3. If the circuit current is unknown, start by using the 10 A and LO terminals.
4. If the measurement is expected to be below 200 mA, connect the test leads to the
200 mA and LO terminals only and remove any leads in the 10 A terminal. See
Figure 3-8.
5. For measurements expected to be 200 mA to 10 A, connect the test leads to the
10 A and LO terminals only. See Figure 3-9.
6. Press B to measure ac current or press E to measure dc current.
7. Apply power to the circuit under test.
The Meter selects the appropriate range in autorange mode. The function and
measurement are displayed.
47


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8808A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8808A in the language / languages: English as an attachment in your email.

The manual is 1,96 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8808A

Fluke 8808A User Manual - German - 102 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info