Introduction and Specifications
Electrical Specifications 1
1-13
Frequency
Gate Time .......................................................... 131 ms
Measurement Method ....................................... AC-coupled input using the ac voltage measurement function.
Settling Considerations ................................... When measuring frequency after a dc offset voltage change, errors
may occur. For the most accurate measurement, wait up to 1 second
to allow input blocking RC time constant to settle.
Measurement Considerations ......................... To minimize measurement errors, shield inputs from external noise
when measuring low voltage, low frequency signals.
Accuracy
Uncertainty
90 days 1 year
Range Frequency
23 °C ± 5 °C 23 °C ± 5 °C
Temperature
Coefficient/°C
Outside 18 – 28 °C
20 Hz – 2 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
2 kHz – 20 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
20 kHz – 200 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
100 mV to
750 V[1,2]
200 kHz – 1 MHz 0.01 + 0.004 0.01 + 0.006 0.002 + 0.002
Notes:
[1] Input > 100 mV
[2] Limited to 8* 107 V Hz
Continuity
Continuity Threshold........................................ 20 Ω
Test Currents .................................................... 1 mA
Response Time ................................................ 100 samples/sec with audible tone
Rate .................................................................... Fast
Maximum Reading ............................................ 199.99 Ω
Resolution ......................................................... 0.01 Ω
Diode Test
Response Time ................................................. 100 samples/sec with audible tone
Rate .................................................................... Fast
Maximum Reading ............................................ 1.9999 V
Resolution ......................................................... 0.1 mV
23


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8808A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8808A in the language / languages: English as an attachment in your email.

The manual is 1,96 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8808A

Fluke 8808A User Manual - German - 102 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info