763940
19
Zoom out
Zoom in
Previous page
1/100
Next page
Introduction and Specifications
Electrical Specifications 1
1-9
Overrange.......................................................... 10 % on the largest ranges of all functions except continuity and diode
test
Remote Interfaces
RS-232C
Warranty
One year
Electrical Specifications
Specifications are valid for 5-½ digit mode and after at least a half-hour warm-up.
DC Voltage Specifications
Maximum Input ................................................. 1000 V on any range
Common Mode Rejection................................. 120 dB at 50 or 60 Hz ±0.1 (1 kΩ unbalance)
Normal Mode Rejection.................................... 80 dB at Slow Rate
A/D Nonlinearity................................................ 15 ppm of range
Input Bias Current ............................................ <30 pA at 25 °C
Settling Considerations ................................... Measurement settling times are affected by source impedance, cable
dielectric characteristics, and input signal changes
Input Characteristics
Resolution
Range Full-Scale
(5-1/2 Digits) Slow Medium Fast
Input Impedance
200 mV 199.999 mV 1 μV 10 μV 10 μV >10 GΩ[1]
2 V 1.99999 V 10 μV 100 μV 100 μV >10 GΩ[1]
20 V 19.9999 V 100 μV 1000 μV 1000 μV 10 MΩ±1 %
200 V 199.999 V 1 mV 10 mV 10 mV 10 MΩ±1 %
1000 V 1000.00 V 10 mV 100 mV 100 mV 10 MΩ±1 %
Notes:
[1] At some dual display measurements, the input impedance of 200 mV and 2 V ranges may be changed to 10 MΩ.
Accuracy
Uncertainty [1]
90 days 1 year
Range
23 °C ± 5 °C 23 °C ± 5°C
Temperature Coefficient/°C
Outside 18 – 28 °C
200 mV 0.01 + 0.003 0.015 + 0.004 0.0015 + 0.0005
2 V 0.01 + 0.002 0.015 + 0.003 0.001 + 0.0005
20 V 0.01 + 0.003 0.015 + 0.004 0.0020 + 0.0005
200 V 0.01 + 0.002 0.015 + 0.003 0.0015 + 0.0005
1000 V 0.01 + 0.002 0.015 + 0.003 0.0015 + 0.0005
Notes:
[1] Uncertainty given as ± (% of reading + % of range)
19


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8808A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8808A in the language / languages: English as an attachment in your email.

The manual is 1,96 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8808A

Fluke 8808A User Manual - German - 102 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info