Detailed Specifications
111
+DC Current Source
Range/Mode Resolution % of Output + Floor
1 Year 2 Year
0.100 to 22.000 mA 1 μA 0.01 % + 3 μA 0.02 % + 3 μA
Temperature Coefficient: 3 % of specified accuracy / °C (<18 °C or >28 °C)
Source mA Compliance Voltage: 18 V maximum
Source mA Open Circuit Voltage: 30 V maximum
+DC Current Simulate (External Loop Power)
Range/Mode Resolution % of Output + Floor
1 Year 2 Year
0.100 to 22.000 mA (Current Sink) 1 μA 0.02 % + 7 μA 0.04 % + 7 μA
Simulate mA Input Voltage: 15 to 50 V dc, add 300 μA to floor when >25 V is present on the loop
Temperature Coefficient: 3 % of specified accuracy / °C (<18 °C or >28 °C)
Resistance Sourcing
Range Resolution % of Output + Floor Allowable Excitation
Current
1 Year 2 Year
10.000 Ω 0.001 Ω 0.01 % + 0.010 Ω 0.015 % + 0.015 Ω 0.1 mA to 10 mA
100.00 Ω [1] 0.01 Ω 0.01 % + 0.02 Ω 0.015 % + 0.03 Ω 0.1 mA to 10 mA
1.0000 kΩ [2] 0.1 Ω 0.02 % + 0.0002 kΩ 0.03 % + 0.0003 kΩ 0.01 mA to 1.0 mA
10.000 kΩ 1 Ω 0.02 % + 0.003 kΩ 0.03 % + 0.005 kΩ 0.01 mA to 1.0 mA
Temperature Coefficient: (0.01 % of output +0.02 % of range / °C (<18 °C or >28 °C).
When connected to mains, accuracy is not specified with conducted RF >1V, 8 to 15 MHz.
[1] Add 0.01 Ω when the excitation current is <1 mA.
[2] Add 0.0015 kΩ when the excitation current is <0.1 mA.
123


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 754 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 754 in the language / languages: English as an attachment in your email.

The manual is 5,57 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 754

Fluke 754 User Manual - German - 130 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info