764336
89
Zoom out
Zoom in
Previous page
1/130
Next page
Documenting Process Calibrator
Process Instrument Calibration
77
To do a pressure switch test: The switch in this example
sets at a high limit of 10 psi. The set condition is a closed
switch contact. For pressure switches, use the Manual Test
selection. For switches where sourcing pressure is not
necessary, use the Auto Test selection to do the test.
1. Connect the test leads between the pressure switch
contact output and the mA Ω RTD (center) jacks on the
Product.
2. Connect the pressure module to the Product, and
connect a pressure line to the switch. Keep the
pressure line vented to atmosphere.
3. If necessary, push for MEASURE mode.
4. Push q for the continuity measure function.
5. Push for SOURCE mode.
6. Push p for the pressure source function.
7. Push to zero the pressure module.
8. Push .
9. Push the As Found softkey.
10. Highlight 1 Pt. Switch Test from the menu and push
.
11. Push to modify the parameters for Setpoint 1.
12. Make these selections:
Setpoint 1 = 10.000 psi
Setpoint Type = High
Set State = Short
13. Push the Done softkey.
14. Set the Tolerance to 0.5 psi.
15. The next parameters, Deadband Min and Deadband
Max, are optional. Do not set them in this example.
These parameters would describe the minimum
allowable size of the deadband.
89


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 754 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 754 in the language / languages: English as an attachment in your email.

The manual is 5,57 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 754

Fluke 754 User Manual - German - 130 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info