764273
20
Zoom out
Zoom in
Previous page
1/50
Next page
709/709H/710
Users Manual
14
Automatic Step and Ramp
To use the automatic step and automatic ramp functions:
1. Use the Main menu to set the Product to source or simulate
current.
2. The Product has separate auto ramp and step times. Use the
menu to set the ramp or step time.
3. Push once to continually step the output from 0 % of span
to 100 % of span and back in increments of 25 % of span at
the specified interval.
4. Push once more to go to auto ramp.
5. Push one of the percent keys, and twice, to turn off the
auto step and ramp.
6. Push twice to continually ramp the output from 0 % of
span to 100 % of span over the specified interval, and then
back over the specified interval.
7. Push one of the percent keys or once more, to turn off the
auto step and ramp.
Valve Test
The Valve Test function turns on or off the valve test feature. See
Figure 13.
Figure 13. Valve Test Enable Screen
The valve test verifies correct operation of valves. In valve test, the
output can be stepped to these values:
3.8 mA
4.0 mA
4.2 mA
8.0 mA
12.0 mA
16.0 mA
19.8 mA
20.0 mA
20.2 mA
The valve test mA values are not affected by the mA span setting.
1. Use the Main menu to set the Product to source or simulate
current.
2. If valve test is not enabled, use the menu to enable it.
3. Push or to step the output to verify the proper valve
operation.
4. Use the menu to disable valve test when done.
20


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 709 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 709 in the language / languages: English as an attachment in your email.

The manual is 5,6 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 709

Fluke 709 User Manual - German - 50 pages

Fluke 709 User Manual - Dutch - 50 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info