764276
4
Zoom out
Zoom in
Previous page
1/16
Next page
English Instruction SheetPanel 4
Using the mA Sourcing (Output) Modes
The Calibrator outputs current for calibrating and testing 0 to 20 mA
and 4 to 20 mA current loops and instruments.
In SOURCE mode, the Calibrator supplies the current.
In SIMULATE mode, the Calibrator simulates a 2-wire transmitter in
an externally-powered current loop.
Changing the mA Output Span
The Calibrator has two mA output spans:
4 mA to 20 mA (0 % to 100 %) [default]
0 mA to 20 mA (0 % to 100 %) [optional]
To change the output span, turn the Calibrator off. Press A + D
simultaneously. The selected setting is saved until it is changed
again.
Sourcing mA
Use SOURCE mode to supply current to a passive circuit.
A path must exist for current to flow between the + and COM
terminals. Otherwise the display indicates an overload (OL) when
you set an output value.
COM
+
707
LOOP CALIBRATOR
30V
CAT
Press until SOURCE
mA is displayed.
Press to step current
up/down in 25 %
steps of full scale.
Turn to step up/down
1 A. Press in and
turn to step up/down
in 100 A increments.
Press simultaneously
once for slow ramp
twice for fast ramp
three times for
stepped ramp.
ADA04F.EPS
4


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 707 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 707 in the language / languages: English as an attachment in your email.

The manual is 0,48 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 707

Fluke 707 User Manual - German - 16 pages

Fluke 707 User Manual - Dutch - 16 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info