Message codes
Table 5 is a list of all message codes that show on the display with InFo or Error.
Table 5. Message Codes
Code Cause Remedy
156 Transverse tilt greater than 10 ° Hold the Meter without a transverse tilt.
162 Calibration mistake Make sure that the device is on a horizontal and flat surface. Do the
calibration procedure again. If the code continues, contact Fluke.
204 Calculation error Do the measurement again.
252 Temperature too high Let the Meter cool down.
253 Temperature too low Let the Meter warm up.
255 Received signal too low,
measurement time too long Change target surface (for example, white paper).
256 Received signal too high Change target surface (for example, white paper)
257 Too much background light Darken target surface.
258 Measurement outside of
measurement range Correct the range.
260 Laser beam interrupted Do the measurement again.
Error Hardware error Turn on and turn off the device 2 to 3 times. If the symbol stays on the
display, then your Meter is defective, contact Fluke.
34


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 414D at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 414D in the language / languages: English as an attachment in your email.

The manual is 2,17 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 414D

Fluke 414D User Manual - German - 38 pages

Fluke 414D User Manual - Dutch - 40 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info