775825
94
Zoom out
Zoom in
Previous page
1/98
Next page
190 Series III
Users Manual
88
7. Adjust the trimmer screw in the probe housing until a pure square wave shows on the display.
For instructions to access the trimmer screw in the probe housing, see the instruction sheet
for the probe.
8. Press 4 to continue with DC calibration. Automatic DC calibration is only possible for
10:1 voltage probes.
The test tool automatically calibrates itself to the probe. During calibration you should not
touch the probe. A message indicates when the DC calibration has completed successfully.
9. Press 4 to exit.
Repeat the procedure for the blue 10:1 voltage probe on input B, the gray 10:1 voltage probe on
input C and the green 10:1 voltage probe on input D.
Note
When using 100:1 voltage probes, choose 100:1 attenuation to perform an adjustment.
100:1 voltage probes require DC-calibration. Trimmer adjustment is not available.
Version and Calibration Information
You can display version number and calibration date:
1. Press U to show the USER key labels.
2. Press 3 to open the VERSION & CALIBRATION screen.
3. Press 4 to close the screen.
The screen gives you information about the model number with software version, the serial
number, the calibration number with latest calibration date, and installed (software) options. The
LICENSE INFO key opens a screen with information about the Open Source Software license.
The test tool specifications are based on a 1 year calibration cycle. Recalibration must be carried
out by qualified personnel. Contact your local Fluke representative for recalibration.
94


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 190-062-III at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 190-062-III in the language / languages: English as an attachment in your email.

The manual is 9.23 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info