ScopeMeter® Test Tool
Recorder Functions
35
4. Press 2 to open the RECORDER OPTIONS menu.
5. Use the cursor and E to highlight the Mode field and select:
a. on Trigger (190-xx4) opens the START SINGLE SWEEP ON TRIGGERING menu
b. on Ext. (190-xx2) opens the START SINGLE SWEEP ON EXT. menu
6. Use the cursor and E to select one of the Conditions: and accept the selection.
For external triggering (190-xx2) continue with:
7. Use the cursor and E to select the trigger slope (Slope) and level (Level).
8. Use the cursor and E to select the 0.12V or 1.2V trigger level and accept all recorder
options.
9. Apply a trigger signal to the red and black ext. trigger banana inputs.
During recording samples are continuously saved in deep memory. The last twelve recorded
divisions show on the screen. See Figure 19. Use View All to display the full memory contents.
Note
To learn more about the Single Shot trigger function, see Waveform Triggers.
Figure 19. Triggered Single Sweep Recording
TrendPlot or Scope Record Analysis
From a TrendPlot or Scope Record you can use the analysis functions CURSORS and ZOOM to
perform detailed waveform analysis. See Replay, Zoom and Cursors.
41


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 190-062-III at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 190-062-III in the language / languages: English as an attachment in your email.

The manual is 9.23 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info