763946
15
Zoom out
Zoom in
Previous page
1/22
Next page
True-rms Multimeter
Product Familiarization
11
Detecting AC Voltage Presence (117)
To detect the presence of ac voltage, place the top of the Meter close to a conductor. The Meter gives an audible
as well as visual indication when voltage is detected. The sensitivity settings are:
Lo: use on flush mounted wall sockets, power strips, flush mounted industrial outlets and various power
cords.
Hi: for ac voltage detection on other styles of recessed power connectors or sockets where the actual ac
voltage is recessed within the connector itself.
The VoltAlert detector works in bare wire applications with voltages as low as 24 V in the Hi setting.
XW Warning
If there is no indication, voltage could still be present. Do not rely on the VoltAlert detector with
shielded wire. Operation may be effected by differences in socket design, insulation thickness
and type.
Making Low Impedance Capacitance Measurements (115, 117)
For making capacitance measurements on cables with ghost voltage:
1. Hold q as you turn on the Meter to enable the low-input impedance Capacitance mode.
2. Wait until LCAP shows on the display.
In this mode, capacitance measurements will have a lower accuracy and lower dynamic range.
Note
This setting is not saved when the Meter is turned off or goes into sleep mode.
VoltAlert
VoltAlert
Lo
RANGE
VoltAlert
Hot Neutral
15


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 114 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 114 in the language / languages: English as an attachment in your email.

The manual is 1,21 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 114

Fluke 114 User Manual - German - 22 pages

Fluke 114 User Manual - Dutch - 22 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info