Technical data26
Image presentation
Automatic image adjustment, type Standard or histogram based on the image content
Manual image adjustment
Linear based, possible to adjust level/span/max./
min.
Image presentation modes
Image modes Thermal, thermal MSX, picture in picture, digital
camera
Infrared image Full color infrared image
Visual image Full color visual image
Multi Spectral Dynamic Imaging (MSX) Thermal image with enhanced detail presentation
Picture in Picture Resizable and movable infrared area on the visual
image
Gallery
Review thumbnail/full image on the camera
Edit measurements/palettes/image modes on
the camera
Measurement
Object temperature range
–40 to +150°C (–40 to +302°F)
0 to +650°C (+32 to +1202°F)
+300 to +2000°C (+572 to +3632°F)
Accuracy
±1°C (±1.8°F) or ±1% @ 25°C (77°F) for tem-
peratures between 5 and 150°C (41 and 302°
F), in range –40 to 150°C (–40 to 302°F)
±2°C (±3.6°F) or ±2% of reading @ 25°C (77°
F) for temperatures between –40 and 2000°C
(–40 and 3632°F)
NOTE
For HSI use, above 30 Hz frame rate, the typical
accuracy will be ±2.5°C (±4.5°F), or 2.5% of
reading @ 25°C (77°F).
Measurement analysis
Spotmeter
10
Area
5 + 5 areas (boxes and circles) with max./min./
average
Profile 1 line profile with max./min. temperature
Automatic hot/cold detection
Auto hot or cold spotmeter markers within the area
and profile
Measurement presets No measurements, Center spot, Hot spot, Cold
spot, User preset 1, User preset 2
User presets The user can select and combine measurements
from any number of spots/boxes/circles/profiles/
delta
Difference temperature Delta temperature between the measurement func-
tions and the reference temperature
Reference temperature Manually set using the difference temperature
Atmospheric transmission correction Automatic, based on the inputs for distance, at-
mospheric temperature, and relative humidity
Optics transmission correction Automatic, based on signals from internal sensors
#T559954; r. AL/37426/37426; en-US
167
181


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Flir T1020 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Flir T1020 in the language / languages: English as an attachment in your email.

The manual is 29,02 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info