Display Indicators
The Primary DisplayIndicate the current function testing values
The Secondary Display It shows the output DCV while you test the insulation
resistance, and the battery voltage while the ACV
The Analog Barindicate the current function testing value in synchronous with the
primary display.
:While testing the insulation resistance, the symbol flashes frequently if the voltage is
over 30V.
•))) : While testing the insulation resistance, the symbol flashes frequently and the
buzzer warns continually if the outside voltage is over 30V.
The symbol •))) “ is indicated while LOΩ35Ω and the BZ warns continuously.
LOCKPush down the LOCK button while you test the insulation resistance and the
symbol is indicated.
LOBATThe display shows LOBAT when the voltage drops below 7.5V
MAX/MIN:Stand for the maximum or the minimum.
ZERO : Digital zero adjusting
HOLD :The digital holding function for the primary display.
ACDC:The indicator for the voltage property.
VMΩΩ: The measured dimension units
Insulation resistance measurements
a)Turn the function switch from the OFF position to the left4000 MΩ/1000V---4000/
MΩ500V---4000 MΩ/250V---1000 MΩ/125V,and chose one of the voltage-block( there
are 4 ranges namely, 4 MΩ40 MΩ400 MΩ4000 MΩ, can be switched automatically for
every voltage-block.)
b) Connect two testing lines to the tested
c)Push down and hold the TEST button /or press the LOCK” keystoke first and then the
TEST button, if the tested is electriferous and its voltage ( AC/DC) is over 30V, it will
refuse work and no high-voltage testing occurs, simultaneity, it shows >30V on the LCD,
the symbol flashes, and the buzzer warns frequently. if the tested is diselectriferous
or its voltage is lower than 30V, it will enter into the formal testing process and brings the
high-voltage. on the primary display, the insulation resistance in MΩ is indicated in-phase
with analog bar; on the secondary display, the tested insulation voltage in V (DC) is
indicated, the symbol flashes and the buzzer warns frequently
25


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Elma DT5500A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Elma DT5500A in the language / languages: English, Danish, Swedish, Norwegian as an attachment in your email.

The manual is 1,17 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info