668438
52
Zoom out
Zoom in
Previous page
1/56
Next page
Elma BM878
side 52
General specification
Display: 3-5/6 digits 6,000 counts
Polarity: Automatic
Update Rate: 5 per second nominal
61 Segments Bar graph: 40 per second max
Operating Temperature: -10C to 40C
Relative Humidity: Maximum relative humidity 90% for temperature up to 28C
decreasing linearly to 50% relative humidity at 40C
Pollution Degree: 2
IP Rating: IP40
Storage Temperature: -20C to 60C, < 80% R.H. (with battery removed)
Altitude: Operating below 2000m
Temperature Coefficient: nominal 0.15 x (specified accuracy)/ C @(-10C ~ 18C or
28C ~ 40C), or otherwise specified
Sensing: AC, True RMS
Safety: Double insulation per IEC/UL/EN61010-1 Ed. 3.0, IEC/EN61010-2-030 Ed. 1.0,
IEC/EN61010-2-033 Ed. 1.0, IEC/UL/EN61010-031 Ed. 1.1 and CAN/CSA-C22.2 No.
61010-1-12 Ed. 3.0 to Category III 1000 V AC & DC and Category IV 600V AC & DC
Compliance to IEC/EN61557:2007 (Per CE requirements, not certified by UL or ETL):
IEC/EN61557-1, IEC/EN61557-2 & IEC/EN61557-4 where applicable
Overload Protections:
Insulation Resistance & mA (Model 878 only): 0.4A/1KV, IR 30kA or better
Earth Continuity Test (Models 878 & 877 only): 0.25A/1KV, IR 30kA or better
V: 1100Vrms
mV, & Others: 1000 Vrms
Transient Protection: 8kV (1.2/50s surge)
E.M.C.: Meets EN61326-1:2006 (EN55022, EN61000-3-2, EN61000-3-3, EN61000-4-2,
EN61000-4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6, EN61000-4-8, EN61000-4-11)
In an RF field of 3V/m:
Total Accuracy = Specified Accuracy + 25 digits
Performance above 3V/m is not specified
Power Supply: Four Alkaline AA batteries (IEC LR6)
Power Consumption: 4.5mA typical except the followings:
ACV
+Hz
& VFD ACV
+Hz
:
7.0mA
Earth Continuity Test:
110mA @20 Range
220mA @2.0 Range
Tester can perform at least 3000 Earth Continuity Test measurements with new alkaline
batteries at room temperature. These are standard tests of 1
with a duty cycle of 5
seconds on and 25 seconds off.
Insulation Resistance @1mA test current:
50V output voltage: 25mA
100V output voltage: 45mA
250V output voltage: 85mA
500V output voltage: 170mA
52


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Elma BM878 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Elma BM878 in the language / languages: English, Danish, Swedish, Norwegian as an attachment in your email.

The manual is 8,92 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info