OPERATION
CAUTION
Before and after hazardous voltage measurements, test the voltage function on a known
source such as line voltage to determine proper meter functioning.
VFD ACV
+Hz
, VFD Hz
+ACV
functions (Model 869 only)
Press the SELECT button momentarily to toggle between the subject functions. Last
selection will be saved as power up default for repeat measurement convenience. By
default, voltage is always set at manual-range 500V to best cope with most Variable
Frequency Drives (VFD) measurements. Press the RANGE button momentarily to select
other ranges only when needed. High noise-rejection frequency measurement algorithm
and Low-pass filter circuit are permanently bundled with all the voltage and frequency
function-ranges within this rotary-switch position.
Note: Hz Input sensitivity varies automatically with voltage (current function alike) range
selected. 5V range has the highest and the 1000V range has the lowest. This VFD
function defaults at the most appropriate trigger level for Variable Frequency Drive
measurements. You can also press the RANGE button momentarily to select another
trigger level (voltage range) manually. If the Hz reading becomes unstable, select higher
voltage range to avoid electrical noise. If the reading shows zero, select lower voltage
range for better sensitivity.
39


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Elma BM 869 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Elma BM 869 in the language / languages: English, Danish, Swedish, Norwegian as an attachment in your email.

The manual is 7,55 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info