652861
8
Zoom out
Zoom in
Previous page
1/40
Next page
8
3.8.6
3.8.7
3.8.8
3.8.9
3.9.0
3.9.1
3.8.1
3.8.2
3.8.3
3.8.4
3.8.5
For using the special programmes for different parts of the body, the DITTMANN
textile electrodes and/or the back pain relief belt, knee pain relief cuff or neck cuff
available as accessories are also ideally suited for a convenient application using
the TENS/EMS device TEN 250 (e. g. for the area H, the stimulation glove HFE
322; for the area F, the stimulation sock SFE 323; for the area E, the elbow
electrode EFE 361; for the area K, the knee electrode KFT 362 or the knee pain
relief cuff KMT 285; for the area B, the back pain relief belt RFT 363 or RGT
284 and for the area N, the neck cuff TNM 275).
When using these products together with the TENS/EMS device TEN 250, please
always follow the operating instructions of these additional products and, above
all, the respective safety instructions!
Avoid any contact with the adhesive electrodes during the therapy. Under certain
circumstances, touching the electrodes results in a short circuit involving an
excessive current density. This excessive current density might cause burns and
injuries!
Please remove the protective foil before applying the adhesive electrodes. The
adhesive strength of the electrodes depends on the condition of the skin, storage
and number of therapy sessions. If the adhesive electrodes do no longer stick
completely on the skin surface, they must be replaced by new adhesive
electrodes. The adhesive electrodes must lie completely flat against the skin in
order to avoid locally high current densities which might result in skin burns. After
the application has been completed, stick the adhesive electrodes back on the
protective foil and keep the adhesive electrodes in the polybag in such a manner
that they cannot dry out. Thus, the adhesion is ensured for longer.
Each person reacts differently to an electric nerve stimulation. If the therapy is
not successful, please seek medical advice.
The adhesive electrodes can be connected to and disconnected from the device by
means of a plug connection with an electric cable.
Avoid pulling the cables directly when removing the plug connections from the
adhesive electrodes in order to avoid damage to the cables. . Only handle the
plugs in order to connect or disconnect the adhesive electrodes!
Only connect the cables and the adhesive electrodes if the TENS/EMS device has
been switched off.
Before using the adhesive electrodes, the areas of the skin where the adhesive
electrodes will be put on must be thoroughly cleaned and dried. The areas of the
skin should be greaseless and clean.
Using the TENS/EMS device may result in skin irritations under certain
circumstances. If skin irritations, e.g. redness, blisters or itching, occur, the
TENS/EMS device should no longer be used! Do not use the adhesive electrodes
permanently on the same part of the body, as this can lead to skin irritations.
If you wish to reposition the adhesive electrodes of the TENS/EMS device during
the therapy, make sure that you switch off the device first.
The adhesive electrodes may only be connected to the TENS/EMS device TEN 250.
Please ensure that the device is switched off when putting on or removing the
adhesive electrodes.
3.8 Usage of the TENS/EMS device
SAFETY INSTRUCTIONS
GB
8


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Dittmann TEN 250 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Dittmann TEN 250 in the language / languages: English as an attachment in your email.

The manual is 1,58 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Dittmann TEN 250

Dittmann TEN 250 User Manual - German - 40 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info