INSTRUCTIONS
Congratulations on your purchase of the CRESTA TH120 body thermometer. It shows your
confidence in our brand and our products. Please be assured you have bought a high
quality product which you will enjoy using for many years to come.
If you require technical support and or have questions about the functioning of the product
please contact our After Sales Service
+ 31 900 4000030
or send an e-mail to
service@sabrenl.com
Read this manual thoroughly before first use, even if you are familiar with this type of product.
It gives you advise how to optimize the use of this product.
READ THESE INSTRUCTIONS CAREFULLY AND FOLLOW UP
This thermometer is intended for the measurement of the human body temperature.
OPERATING INSTRUCTIONS
Before using, please disinfect the probe first. To switch on press the ON/OFF button next to the
display; a short beep will sound, indicating that the thermometer is operational. At the same
time the thermometer runs a self-check test, during which all digital segments appear on the
LDC display. When the letters “Lo” and a flashing “C” display, the thermometer is ready for use.
If the ambient temperature is below 32°C, then “Lo°C” will appear on the LCD display and if
the temperature is higher then 42°C then “Hi°C” will appear on the LCD display. During the
reading the current temperature is displayed continuously and the “C” symbol flashes. The
measurement is completed when a constant temperature value has been reached. The
temperature value is considered constant when the temperature rises less than 0,1°C within 16
seconds. As soon as a constant temperature value is reached, a beep will sound 4 times and
the “C” symbol will stop flashing. The highest measured temperature appears on the LCD
display. Please note however that this thermometer is a maximum thermometer, i.e. the
displayed temperature can increase slightly if measurement continues after the beep. This is
particularly the case with axillary measurements, the temperature value should be recorded
which approximates the core body temperature. In this instance please note the description
under “Methods of measuring temperature”. When the measurement is completed, please
switch off by pressing the ON/OFF button. After the temperature has been displayed the
thermometer will shut off automatically after 10 minutes.
MEMORY FUNCTION
Turn on the thermometer, the last measured temperature will display on the LCD display for
approximately 2 seconds after the self-check. The reading will be overwritten after the LCD
shows “37.0°C”, i.e. the thermometer is ready for a next measurement, whether the new
measurement is complete or not.