POWER SUPPLY
Li_ion rechargeable battery (3.7 V, 2000mAh)
The rechargeable Li-ion battery of Compex® Wired 3.0 has an expected life of more than 300
full charge cycles.
Power Supply Adapter:
Manufacturer name: Shenzhen Huoniu Technology Co., Ltd.
Model #: HNEM050200UU
Input: 100-240 VAC, 50/60 Hz, 0.35A Max
Output: 5.0 VDC, 2.0 A
OUTPUT SPECIFICATIONS
Pulse Shape: Constant rectangular current with pulse compensation to eliminate any direct current
component to prevent residual polarization at skin level.
Maximum Pulse Intensity: 120 mA
Maximum Pulse Voltage: 120 V
Pulse Intensity Increments: manual adjustment of stimulation intensity from 0 to 999 (energy) in
minimum increments of 0.5 mA.
Pulse Width: 200 to 400 μs
Maximum Electrical Charge Per Pulse: 96 microcoulombs (2x48 μC, compensated). Standard
pulse ramp-up time: 3 μs (20%-80% of maximum current).
Pulse Frequency: 1 to 120 Hz
Note: the electrodes may deliver an output of more than 10mA or 10V averaged when you choose
high intensity levels
DESCRIPTION OF ACCESSORIES
For a list of detailed accessories please visit www.compex.com
Four black snap-connection stimulation cables:
Device connector: 6-pin; Electrode connector: female snap; Length: 1500 mm
TABLE 1: RECOMMENDATIONS AND DECLARATION BY THE MANUFACTURER CONCERNING
ELECTROMAGNETIC EMISSIONS
The
Compex® Wired 3.0
is intended for use in the electromagnetic environment specified below.
The customer or the user of the Compex® Wired 3.0 should assure that it is used in such an environment.
Self-Adhesive Electrodes:
4 small electrodes (5 x 5 cm, 2 x 2 inch)
2 large electrodes (5 x 10 cm, 2 x 4 inch)
Compex USB Charging Cable
Emission Tests Compliance Electromagnetic Environment - Guidance
CISPR 11
RF Emissions Group 1
The Compex® Wired 3.0 uses RF energy only for its internal
operation. Consequently, its RF emissions are very low and are
unlikely to interfere with any adjacent electrical device.
CISPR 11
RF Emissions Class B
The Compex® Wired 3.0 is suitable for use in any
establishment, including a private dwelling and a place
connected directly to the low voltage mains supply which
powers residential buildings.
Harmonic Emissions
IEC 61000-3-2 Not Applicable
Voltage Fluctuations/
Emission Oscillations
IEC 61000-3-3
Not Applicable
TABLES 2 & 3: RECOMMENDATIONS AND DECLARATION BY THE MANUFACTURER -
ELECTROMAGNETIC IMMUNITY
The Compex® Wired 3.0 is designed for use in the electromagnetic environment stipulated
below. The customer or the user of the Compex® Wired 3.0 must ensure that it is used in this
recommended environment.
Note: UT is the a.c. mains voltage prior to application of the test level.
Immunity Test IEC 60601 Test
Level Compliance Level Electromagnetic
Environment - Guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
±6 k V a t t he c ont act
±8 kV in air
±6 kV at the contact
±8 kV in air
Floors must be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at a minimum of 30%.
Electrical fast
transient/ burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
Not Applicable
System battery-
powered
The quality of the power
supply should be that of a
typical commercial or hospital
environment.
Surge (1)
IEC 61000-4-5
±1 kV differential
mode
±2 kV common mode
Not Applicable
System battery-
powered
The quality of the power
supply should be that of a
typical commercial or hospital
environment.
Power frequency
(50/60Hz)
magnetic field IEC
61000-4-8
3 A/m
Magnetic fields at the mains
frequency should be at the level
of a representative site located
in a typical commercial or
hospital environment.
18


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Compex Wired 3.0 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Compex Wired 3.0 in the language / languages: English as an attachment in your email.

The manual is 2.11 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info