Figure 4-3
A collimated telescope
should appear
symmetrical with the
central obstruction
centered in the star's
diffraction pattern.
Before you begin the collimation process, be sure that your telescope is in thermal equilibrium with the
surroundings. Allow 45 minutes for the telescope to reach equilibrium if you move it between large temperature
extremes.
To verify collimation, view a star near the zenith. Use a medium to high power ocular 12mm to 6mm focal
length. It is important to center a star in the center of the field to judge collimation. Slowly cross in and out of
focus and judge the symmetry of the star. If you see a systematic skewing of the star to one side, then re-collimation
is needed.
To accomplish this, you need to tighten the secondary collimation screw(s) that move the star across the field toward
the direction of the skewed light. These screws are located in the secondary mirror holder (see figure 4-1). Make
only small 1/6 to 1/8 adjustments to the collimation screws and re-center the star by moving the scope before
making any improvements or before making further adjustments.
To make collimation a simple procedure, follow these easy steps:
1. While looking through a medium to high power eyepiece, de-focus a bright star until a ring pattern with a
dark shadow appears (see figure 4-2). Center the de-focused star and notice in which direction the central
shadow is skewed.
2. Place your finger along the edge of the front cell of the telescope (be careful not to touch the corrector
plate), pointing towards the collimation screws. The shadow of your finger should be visible when looking
into the eyepiece. Rotate your finger around the tube edge until its shadow is seen closest to the narrowest
portion of the rings (i.e. the same direction in which the central shadow is skewed).
3. Locate the collimation screw closest to where your finger is positioned. This will be the collimation screw
you will need to adjust first. (If your finger is positioned exactly between two of the collimation screws,
then you will need to adjust the screw opposite where your finger is located).
4. Use the hand control buttons to move the de-focused star image to the edge of the
field of view, in the same direction that the central obstruction of the star image is
skewed.
5. While looking through the eyepiece, use an Allen wrench to turn the collimation
screw you located in step 2 and 3. Usually a tenth of a turn is enough to notice a
change in collimation. If the star image moves out of the field of view in the
direction that the central shadow is skewed, than you are turning the collimation
screw the wrong way. Turn the screw in the opposite direction, so that the star
image is moving towards the center of the field of view.
6. If while turning you notice that the screws get very loose, then simply tighten the
other two screws by the same amount. Conversely, if the collimation screw gets
too tight, then loosen the other two screws by the same amount.
Figure 4-2 -- Even though the star pattern appears the same on both sides of focus, they are asymmetric. The
dark obstruction is skewed off to the left side of the diffraction pattern indicating poor collimation.
16


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Celestron EdgeHD 9.25 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Celestron EdgeHD 9.25 in the language / languages: English as an attachment in your email.

The manual is 1,72 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info