FI
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
(9.397 pt)
Immunity Test IEC 60601
Test Level Compliance Level Electromagnetic Environment -
Guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Risk assessment on the CEFAR EASY
Electrotherapy System indicates
the compliance levels claimed are
acceptable when ESD-precautionary
measures are taken.
Electrical fast
transient/burst IEC
61000-4-4
±2kV for power supply lines
±1kV for input/output lines
Not Applicable -
Battery powered
Not Applicable
- signal lines less
then 3 meters
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
+ 1kV dierential mode (line to line)
+ 2kV common mode (line to ground)
Not Applicable -
Battery powered
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% UT (>95% dip in UT) for 0,5 cycle
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 25 cycles
<5% UT (>95% dip in UT) for 5 sec
Not Applicable -
Battery powered
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
CEFAR EASY Electrotherapy System
requires continued operation during
power mains interruptions, it is
recommended that the CEFAR EASY
Electrotherapy System be powered
from an uninterrupted power supply.
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
NOTE: UT is the AC mains voltage prior to application of the test level.
180


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Cefar Easy at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Cefar Easy in the language / languages: English, German, Dutch, French, Italian, Portuguese, Swedish, Spanish, Finnish as an attachment in your email.

The manual is 6.24 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info