819996
17
Zoom out
Zoom in
Previous page
1/21
Next page
38
THE STIMULATION DOES NOT FEEL THE SAME AS USUAL
· Check that all settings are correct (see section STEP-BY-STEP-USE) and make sure that the electrodes are
correctly placed.
· Slightly change the position of the electrodes.
THE STIMULATION FEELS UNPLEASANT
· The skin is irritated. For advice on skin care, see chapter PRECAUTIONARY MEASURE.
· The electrodes begin to lose their stickiness and do not stick properly to the skin. Moisten the adhesive
surface with a few drops of water before placing on the skin.
· The electrodes are worn out and need to be replaced.
· There is insufficient electrode gel on the carbon rubber electrodes.
· Slightly change the position of the electrodes.
THE STIMULATION FEELS WEAK OR NOT AT ALL
· Check if the batteries need to be replaced, see chapter REPLACEMENT OF BATTERIES.
· Electrodes are too old and need replacement.
THE BROKEN CIRCUIT SYMBOL IS SHOWN ON THE DISPLAY SYMBOL
The broken circuit symbol indicates that the resistance is too high, or that a cable is broken.
· A too high resistance can be caused by a bad connection between the electrodes and your skin, or that the
electrodes need to be replaced.
· A cable breakage can be checked by pressing the cable´s pins against one another while increasing the
amplitude for the corresponding channel to 11 mA. If the amplitude now drops to 0.0 mA and starts
flashing, the cable needs to be replaced.
Note! Never increase the amplitude above 20 mA when you check for cable breaks, since this can damage
the stimulator.
THE STIMULATOR IS NOT WORKING
If the error symbol appears on the display when you start the stimulator, it means that the stimulator is
broken and needs to be replaced.
Note! Do not use the stimulator – contact your Cefar dealer.
Cefar will only be responsible for service and repairs performed by Cefar or a distributor appointed by
Cefar.
10. TROUBLESHOOTING
17


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Cefar Basic at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Cefar Basic in the language / languages: English as an attachment in your email.

The manual is 2.43 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Cefar Basic

Cefar Basic User Manual - English, German - 256 pages

Cefar Basic User Manual - German - 26 pages

Cefar Basic User Manual - Dutch, French, Italian, Portuguese, Swedish, Spanish, Norwegian - 256 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info