590633
55
Zoom out
Zoom in
Previous page
1/66
Next page
55/66 Device handbook CAM mobile, 06/2014
C4 Operation
C4.1 Measuring current
Read safety section of instructions before operating this product.
Ensure conductor to be tested is de-energised
Connect the output of the electronics to the input of an oscilloscope or other data recording
device.
The flexible current probe is not for use on conductors with a potential of over 1000V.
Wrap the flexible probe around the conductor to be tested, close coupling. Energise the circuit
under test. Locate the coupling away from nearby conductors.
Always use appropriate equipment for personal protection. Do not use the flexible
current probe to measure bare conductors unless wearing protective clothing
suitable for high voltage work.
C4.2 Operation
To activate unit, move the rotary switch from the “off” position to the required measuring range.
If the value of current being measured is unknown, first select the highest range and then
reduce accordingly.
C5 Maintenance
Do not use ACP 3000_5 or ACP 3003_5 if damaged.
Always inspect the integrator unit, connecting cable, and flexible probe for damage before use.
To avoid electric shock, keep the ACP clean and free of surface contamination.
Use Isopropyl alcohol to clean the electronics unit and the probe
Make sure the flexible probe, connecting cable, and electronics enclosure are dry before further
use
55


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Camille Bauer SINEAX CAM mobile at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Camille Bauer SINEAX CAM mobile in the language / languages: English as an attachment in your email.

The manual is 8,92 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info