MAINTENANCE
The NEW ASPIRET suction equipment does not need maintenance or lubrication.
It is necessary to check functioning and instrument before every use.
Unpack the instrument and always check integrity of plastic parts and feeding cable, they might have been
damaged during previous use.
Connect cable to electrical network and turn switch on.
Close the aspiration outlet with your finger and with suction regulator in maximum vacuum position check
that the vacuum indicators reaches -75 kPa (-0.75 bar) maximum.
Rotate the knob from right to left and check the aspiration regulating control.
The vacuum indicator should go down -25 kPa (-0.25 bar).
Verify that loud noises are not present, these can indicate wrong functioning.
A protection fuse (F 1x1.6A 250V and F 1 x 4A 250V for voltage 110V/60Hz) reachable from exterior and it
situated in the plug protects the instrument. For use replacing, always check the type and the range
indicated.
Fault type Cause Solution
1. No aspiration Jar Cap badly screwed
down
Unscrewed the cap, then rescrew it correctly
2. No aspiration Lid seal not in its seat Unscrew the cap and insert the seal properly
in its seat
3. The float doesn’t close If the cap has been
washed, ensure that the
float is not partially
detached
Insert the float into it’s place
4. The float doesn’t close
The float it’s covered by
dirty material
Unscrewed the cap, leave the and put in on
autoclave
5. Low suction Foam inside the jar Fill the jar to 1/3 full of ordinary water
6. No aspiration due to
flow leakage of mucus
Filter blocked Replace filter
Faults 1 - 2 - 3 - 4 - 5 - 6 None of the remedies has
achieved the desired
results
Contact the seller or CA-MI After-sales
Assistance Service
If the overfill security system it’s activated, don’t proceede with the liquid aspiration.
If the overfill security system doesn’t work there are two cases:
1° case – If the overfill security system doesn’t work the aspiration will be stopped by the bacteriological filter
who avoid the liquid penetration inside the device.
2° case – If both the security system doesn’t work, there is the possibility that liquid comes inside the device,
in this case return the device to CA-MI technical service.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING,
PLEASE CONTACT CA-MI TECHNICAL SERVICE.
CA-MI DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL SERVICE
CHECKING, APPEARS TO BE TAMPERED.
14


Need help? Post your question in this forum.

Forumrules
1

Forum

cami-new-aspiret-surgical-aspirator

Reset search

  • Why does my filter repeatedly gets wet? I've just replaced it with a new filter, and now, there's clear fluid in it again. Anybody with this experience? Submitted on 6-2-2020 at 05:04

    Reply Report abuse


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Cami New Aspiret Surgical Aspirator at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Cami New Aspiret Surgical Aspirator in the language / languages: English, German, French, Italian, Spanish as an attachment in your email.

The manual is 0,83 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Cami New Aspiret Surgical Aspirator

Cami New Aspiret Surgical Aspirator User Manual - English - 9 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info