MAINTENANCE
The NEW ASPIRET suction equipment does not need maintenance or lubrication.
It is, however, necessary to inspect the unit before each use. Unpack the instrument and always check the plastic
parts for any damage that may have occurred during prior use. Connect cable to electrical network and turn switch on.
Close the aspiration outlet with your finger and with suction regulator at maximum check that the vacuum position
reaches at least -75 kPa (-0.75 bar) maximum. Rotate the knob from right to left and check the aspiration regulating
control.
The vacuum indicator should go down to -25 kPa (-0.25 bar). Check that no loud noises are present.
A fuse (F 1x1.6A 250V) is located in the plug to protect the instrument. If required, replace with another or the
correct type.
Fault type Cause Solution
1. The suction unit
doesn’t work
Cable is damaged
External power source failure
Replace the cable
Check the external power source
2. No aspiration
Jar Cap badly screwed down Unscrewed the cap, then rescrew it correctly
3. No aspiration Lid seal not in its seat Unscrew the cap and insert the seal properly in its
seat
4. The Vacuum power
on the patient side is either
very low or absent
a) Vacuum regulator set to
minimum
b)Protection filter blocked or
damaged
c) Connection tubes blocked,
kinked or disconnected
d)Shut-off valve blocked or
damaged
e) Pump motor damaged
a) Turn the vacuum regulator clockwise and check the
value of the vacuum on the gauge
b)Replace the filter
c) Replace or reconnect the tubes, check the jar
connections
d)Empty the jar, or disconnect the tube from the jar
and unblock the shut-off valve. The unit twill only
work in the upright position
e) Refer to authorised service personnel
5. The float doesn’t close If the cap has been washed,
ensure that the float is not
partially detached
Insert the float into it’s place
6. The float doesn’t close The float it’s covered by dirty
material
Unscrewed the cap, leave the and put in on autoclave
7. Low suction Foam inside the jar Fill the jar to 1/3 full of ordinary water
Faults 1 - 2 - 3 - 4 - 5 - 6 - 7 None of the remedies has
achieved the desired results
Contact the seller or CA-MI After-sales Assistance
Service
If the overfill security system fails, continued aspiration will be stopped by the bacteriological filter.
Should this back-up security also fail, there is a risk of penetration of liquid into the device. Don’t attempt to proceed
with aspiration under these circumstances.
Return the device to CA-MI for service.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT CA-MI
TECHNICAL SERVICE. THE MANUFACTURER DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL
SERVICE CHECKING, APPEARS TO BE TAMPERED.
7


Need help? Post your question in this forum.

Forumrules
1

Forum

cami-new-aspiret-surgical-aspirator

Reset search

  • Why does my filter repeatedly gets wet? I've just replaced it with a new filter, and now, there's clear fluid in it again. Anybody with this experience? Submitted on 6-2-2020 at 05:04

    Reply Report abuse


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Cami New Aspiret Surgical Aspirator at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Cami New Aspiret Surgical Aspirator in the language / languages: English as an attachment in your email.

The manual is 0,66 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Cami New Aspiret Surgical Aspirator

Cami New Aspiret Surgical Aspirator User Manual - English, German, French, Italian, Spanish - 40 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info